Experimental evaluation of direct thromboaspiration efficacy according to the angle of interaction between the aspiration catheter and the clot

Published inJournal of NeuroInterventional Surgery, p. neurintsurg-2020-016889
Publication date2021
Abstract

Background: Successful direct thromboaspiration (DTA) is related to several factors such as clot consistency, size, and location. It has also been demonstrated recently that the angle of interaction (AOI) formed by the aspiration catheter and the clot is related to DTA efficacy. The aims of this study were three-fold: (a) to confirm the clinical finding that the AOI formed by the aspiration catheter and the clot influence DTA efficacy; (b) to evaluate to what extent this influence varies according to differences in clot consistency and size; and (c) to validate stent retriever thrombectomy as an effective rescue treatment after DTA failure in the presence of an unfavorable AOI.

Keywords
  • Device
  • Stent
  • Stroke
  • Thrombectomy
Citation (ISO format)
BERNAVA, Gianmarco et al. Experimental evaluation of direct thromboaspiration efficacy according to the angle of interaction between the aspiration catheter and the clot. In: Journal of NeuroInterventional Surgery, 2021, p. neurintsurg–2020–016889. doi: 10.1136/neurintsurg-2020-016889
Main files (1)
Article (Accepted version)
Identifiers
ISSN of the journal1759-8478
242views
106downloads

Technical informations

Creation2021/01/25 10:27:00
First validation2021/01/25 10:27:00
Update time2023/03/16 00:03:51
Status update2023/03/16 00:03:50
Last indexation2024/02/12 12:01:55
All rights reserved by Archive ouverte UNIGE and the University of GenevaunigeBlack