Scientific article
English

Multi-time-lag PIV analysis of steady and pulsatile flows in a sidewall aneurysm

Published inExperiments in Fluids, vol. 55, no. 1746
Publication date2014
Abstract

The effect of inflow waveform on the hemodynamics of a real-size idealized sidewall intracranial aneurysm (IA) model was investigated using particle imaging velocimetry (PIV). For this purpose, we implemented an error analysis based on several PIV measurements with different time lags to ensure high precision of velocity fields measured in both the IA and the parent artery. The relative error measured in the main part of the circulating volume was <1% despite the three orders of magnitude difference of parent artery and IA dome velocities. Moreover, important features involved in IA evolution were potentially emphasized from the qualitative and quantitative flow pattern comparison resulting from steady and unsteady inflows. In particular, the flow transfer in IA and the vortical structure were significantly modified when increasing the number of harmonics for a typical physiological flow, in comparison with quasi-steady conditions.

Citation (ISO format)
BOUILLOT, Pierre et al. Multi-time-lag PIV analysis of steady and pulsatile flows in a sidewall aneurysm. In: Experiments in Fluids, 2014, vol. 55, n° 1746. doi: 10.1007/s00348-014-1746-0
Main files (1)
Article (Published version)
accessLevelRestricted
Identifiers
Additional URL for this publicationhttp://link.springer.com/10.1007/s00348-014-1746-0
Journal ISSN0723-4864
333views
1downloads

Technical informations

Creation19/11/2019 13:49:00
First validation19/11/2019 13:49:00
Update15/03/2023 18:33:28
Status update15/03/2023 18:33:28
Last indexation31/10/2024 17:12:40
All rights reserved by Archive ouverte UNIGE and the University of GenevaunigeBlack