Title

Polymerase-chain reaction/electrospray ionization-mass spectrometry for the detection of bacteria and fungi in bronchoalveolar lavage fluids: a prospective observational study

Authors
Huttner, A
Renzi, G
Published in Clinical Microbiology and Infection. 2014, vol. 20, no. 12, p. O1059-66
Abstract PLEX-ID uses polymerase chain reaction-electrospray ionization/mass spectrometry for rapid identification of infectious agents in clinical samples. We evaluated its concordance with our centre's standard methods (SM) for bacterial and fungal detection in bronchoalveolar lavage (BAL) fluid in a prospective observational cohort study. The primary outcome was concordance (%) between SM and PLEX-ID. Secondary outcomes included concordance when excluding commensal oral flora, detection of resistance genes, and PLEX-ID's potential impact on clinical management, as determined by two independent reviewers. Included were 101 specimens from 94 patients. BALs were performed primarily for suspected pneumonia (76/101, 75%) and lung transplant work-ups (12/101, 12%). Most specimens yielded at least one organism by either method (92/101, 91%). Among all microorganisms detected (n = 218), 83% and 17% were bacterial and fungal, respectively. Overall concordance between SM and PLEX-ID was 45% (45/101). Concordance increased to 66% (67/101) when discordance for commensal flora was excluded. PLEX-ID failed to detect 21% of all 183 SM-identified organisms, while SM did not identify 28% of the 191 PLEX-ID-identified organisms (p <0.001). There was low concordance for mecA detection. Two infectious-disease specialists' analyses concluded that in most of the 31 discordant, non-commensal cases, PLEX-ID results would have had little or no impact on patient management; in eight cases, however, PLEX-ID would have led to 'wrong decision-making'. The tested version of PLEX-ID concurred weakly with standard methods in the detection of bacteria and fungi in BAL specimens, and is not likely to be useful as a standalone tool for microbiological diagnosis in suspected respiratory infections.
Identifiers
PMID: 25529027
Full text
Article (Published version) (198 Kb) - document accessible for UNIGE members only Limited access to UNIGE
Structures
Research groups Groupe Harbarth Stephan (Staphylocoques dorés résistants à la méthicilline) (866)
Groupe Laurent Kaiser (virologie clinique) (668)
Analyse génomique et fonctionnelle du staphylocoque doré (604)
Citation
(ISO format)
HUTTNER, A et al. Polymerase-chain reaction/electrospray ionization-mass spectrometry for the detection of bacteria and fungi in bronchoalveolar lavage fluids: a prospective observational study. In: Clinical Microbiology and Infection, 2014, vol. 20, n° 12, p. O1059-66. https://archive-ouverte.unige.ch/unige:73945

129 hits

0 download

Update

Deposited on : 2015-07-07

Export document
Format :
Citation style :