Scientific article
English

Percutaneous treatment of liver tumors with an adapted probe for cooled-tip, impedance-controlled radio-frequency ablation under open-magnet MR guidance: initial results

Published inEuropean radiology, vol. 13, no. 5, p. 1100-1105
Publication date2003
Abstract

Percutaneous radio-frequency (RF) ablation of liver tumors is usually performed under guidance of real-time US, but some tumor nodules in some patients cannot be adequately visualized with this technique. We report our preliminary results with an MR-compatible, internally perfused 17-G RF probe adapted to a standard RF generator for impedance-controlled RF ablation under MR guidance. Following initial testing of the probe for MR compatibility, artifacts and macroscopic effects on an ex vivo pig liver, four patients with eight neoplastic liver nodules (five metastatic and three primary), which could not be properly targeted by US, were treated with the cooled-tip technique under MRI guidance in an open 0.23-T magnet. Metallic artifacts produced by the probe were useful for accurate placement and did not interfere with MRI monitoring at the end of the procedure. Based on imaging findings, the immediate result of RF was considered adequate in all instances. Local recurrence occurred in one instance after 6 months, requiring repeat treatment. No adverse effects were noted. Initial experience suggests that the probe we used allows to perform impedance-controlled cooled-tip RF ablation of liver tumors under open-magnet MR guidance.

Keywords
  • Adult
  • Aged
  • Artifacts
  • Carcinoma, Hepatocellular/diagnosis/pathology/surgery
  • Catheter Ablation/instrumentation
  • Cold Temperature
  • Colorectal Neoplasms/diagnosis/secondary/surgery
  • Contrast Media
  • Electric Impedance
  • Electrocoagulation
  • Equipment Design
  • Female
  • Follow-Up Studies
  • Humans
  • Liver Neoplasms/diagnosis/pathology/surgery
  • Magnetic Resonance Imaging/methods
  • Male
  • Necrosis
  • Neoplasm Recurrence, Local/diagnosis/pathology/surgery
  • Surgery, Computer-Assisted
  • Tomography, X-Ray Computed
  • Treatment Outcome
Citation (ISO format)
KELEKIS, Alexis et al. Percutaneous treatment of liver tumors with an adapted probe for cooled-tip, impedance-controlled radio-frequency ablation under open-magnet MR guidance: initial results. In: European radiology, 2003, vol. 13, n° 5, p. 1100–1105. doi: 10.1007/s00330-003-1847-2
Main files (1)
Article (Published version)
accessLevelRestricted
Identifiers
Journal ISSN0938-7994
558views
1downloads

Technical informations

Creation03/06/2014 17:30:00
First validation03/06/2014 17:30:00
Update time14/03/2023 22:17:44
Status update14/03/2023 22:17:44
Last indexation30/10/2024 18:28:51
All rights reserved by Archive ouverte UNIGE and the University of GenevaunigeBlack