Scientific article
OA Policy
English

Using the Tissue Impulse Response Function to Streamline Fractionated MRgFUS-Induced Hyperthermia

Published inCancers, vol. 17, no. 3, 515
First online date2025-02-04
Abstract

Background/Objectives : Combining radiation therapy with mild hyperthermia, especially via magnetic resonance-guided focused ultrasound (MRgFUS), holds promise for enhancing tumor control and alleviating symptoms in cancer patients. However, current clinical applications of MRgFUS focus primarily on ablative treatments, and using MRI guidance for each radiation session increases treatment costs and logistical demands. This study aimed to test a streamlined workflow for repeated hyperthermia treatments that reduces the need for continuous MRI monitoring, using an approach based on impulse response function (Green's function) to optimize acoustic power settings in advance. Methods : We implemented the Green's function approach in a perfused, tissue-mimicking phantom, conducting 30 experiments to simulate hyperthermia delivery via MRgFUS. Pre-calculated acoustic power settings were applied to maintain a stable hyperthermia target without the need for real-time feedback control from MRI thermometry. Additionally, a retrospective analysis of patient thermometry data from MRgFUS sonications was performed to assess feasibility in clinical contexts.

Results : Our experiments demonstrated consistent, stable hyperthermia (+7 °C) for 15 min across varying perfusion rates, outperforming conventional closed-loop MRI feedback methods in maintaining temperature stability. The retrospective analysis confirmed that this method is noise-robust and clinically applicable.

Conclusions : This off-line approach to hyperthermia control could simplify the integration of MRgFUS hyperthermia in cancer treatment, reducing costs and logistical barriers. These findings suggest that our method may enable the broader adoption of hyperthermia in radiation therapy, supporting its role as a viable adjuvant treatment in oncology.

Keywords
  • HIFU
  • MRI
  • Hyperthermia therapy
Funding
  • Swiss Innovation Agency [103.638 IP-LS]
Citation (ISO format)
GUILLEMIN, Pauline Coralie et al. Using the Tissue Impulse Response Function to Streamline Fractionated MRgFUS-Induced Hyperthermia. In: Cancers, 2025, vol. 17, n° 3, p. 515. doi: 10.3390/cancers17030515
Main files (1)
Article (Published version)
Identifiers
Additional URL for this publicationhttps://www.mdpi.com/2072-6694/17/3/515
Journal ISSN2072-6694
16views
55downloads

Technical informations

Creation06/01/2026 13:50:31
First validation23/01/2026 14:38:18
Update23/01/2026 14:38:18
Status update23/01/2026 14:38:17
Last indexation23/01/2026 14:38:19
All rights reserved by Archive ouverte UNIGE and the University of GenevaunigeBlack