Abstract

Purpose of review: Bioelectrical impedance analysis (BIA) is a widely used, noninvasive method for assessing body composition. Recent technological advances have diversified BIA devices in terms of measurement frequency, electrode configuration, and portability. This review outlines key criteria for selecting a BIA system according to clinical or research needs.

Recent findings: Single-frequency BIA (SF-BIA) devices, typically consumer-grade with hand-to-hand or foot-to-foot configurations, are affordable and easy to use but often lack raw data access, clinical validation, and regulatory certification. In contrast, multifrequency BIA (MF-BIA) systems, especially octopolar models, enable segmental analysis and provide greater accuracy for evaluating fluid distribution and lean mass. However, they are costlier, depend on proprietary algorithms, and generally require standing measurements. In hospital settings, portable MF-BIA devices that allow supine, tetrapolar or octopolar assessments are preferable, particularly for use with bedridden patients. Across all contexts, standardized measurement protocols and access to raw parameters (Z, R, Xc, PhA) are essential to apply accurate, population-specific predictive equations.

Summary: Reliable use of BIA requires careful consideration of device type, data accessibility, and methodological consistency. Portable, regulatory-certified MF-BIA systems with tetrapolar or octopolar configurations and access to raw data offer the most accurate and adaptable solutions for clinical and research applications.

Keywords
  • Bioelectrical impedance analysis
  • Body composition
  • Electrode configuration
  • Nutritional assessment
Citation (ISO format)
DUPERTUIS, Yves Marc et al. Bioelectrical impedance analysis instruments : how do they differ, what do we need for clinical assessment ? In: Current opinion in clinical nutrition and metabolic care, 2025, vol. 28, n° 5, p. 379–387. doi: 10.1097/MCO.0000000000001142
Main files (1)
Article (Published version)
Identifiers
Journal ISSN1363-1950
49views
684downloads

Technical informations

Creation28/07/2025 14:39:40
First validation18/08/2025 09:19:17
Update18/08/2025 09:19:17
Status update18/08/2025 09:19:17
Last indexation18/08/2025 09:19:18
All rights reserved by Archive ouverte UNIGE and the University of GenevaunigeBlack