Abstract

Introduction: The vestibular implant is a neuroprosthesis which offers a potential treatment approach for patients suffering from vestibulopathy. Investigating the influence of electrical stimulation parameters is essential to improve the vestibular implant response. Optimization of the response focuses on the electrically evoked vestibulo-ocular reflex. It aims to facilitate high peak eye velocities and adequate alignment of the eye movement responses. In this study, the basic stimulation parameters of the vestibular implant were tested for their effect on the electrically evoked vestibulo-ocular reflex.

Methods: Four stimulation parameters, including the stimulation amplitude, phase duration, stimulus rate and speed of change of stimulation, were systematically tested in a cohort of nine subjects with a vestibulo-cochlear implant. These parameters were tested to evaluate their effect on fitting settings (i.e., threshold of activation, upper comfortable limit and dynamic range) as well as on the electrically evoked vestibulo-ocular reflex (peak eye velocity and alignment).

Results: It was confirmed that, in addition to current amplitude, the peak eye velocity of the response can be increased by increasing the phase duration and pulse rate. Both parameters have little effect on the alignment of the eye response. However, a longer phase duration decreased the range between the threshold of activation and the upper comfortable limit of the electrical stimulation (i.e., dynamic range). Furthermore, these results show that next to the amplitude of the stimulation, the speed of change in stimulation has a determinative positive effect on the peak eye velocity.

Conclusion: The observations in this study imply that the vestibular implant response, in terms of peak eye velocity, can be optimized with a higher pulse rate and longer phase duration. However, this comes at a trade-off between the dynamic range and power consumption. This study provides essential insights for fitting strategies in future vestibular implant care.

Keywords
  • Electrically evoked vestibulo-ocular reflex
  • Phase duration
  • Pulse rate
  • Stimulation amplitude
  • Vestibular implant
  • Vestibular stimulation
Citation (ISO format)
VAN BOXEL, Stan C J et al. The vestibular implant: effects of stimulation parameters on the electrically-evoked vestibulo-ocular reflex. In: Frontiers in neurology, 2024, vol. 15, p. 1483067. doi: 10.3389/fneur.2024.1483067
Main files (1)
Article (Published version)
Secondary files (7)
Appendix - Figure 1
accessLevelPublic
Appendix - Figure 2
accessLevelPublic
Appendix - Figure 3
accessLevelPublic
Show more
Identifiers
Journal ISSN1664-2295
60views
32downloads

Technical informations

Creation27/11/2024 17:38:02
First validation04/02/2025 08:27:44
Update04/02/2025 08:27:44
Status update04/02/2025 08:27:44
Last indexation04/02/2025 08:27:45
All rights reserved by Archive ouverte UNIGE and the University of GenevaunigeBlack