Mechanical stability of posterior implant-supported monolithic zirconia cantilever on titanium-base abutments. An in vitro study

Publication date

2024-02-28

First online date

2024-02-28

Abstract

Purpose: Investigate survival and technical complications of two-unit posterior implant-supported cantilever made of monolithic zirconia on titanium-base abutments (Zr-TiB) vs. porcelain-fused-to-metal on castable gold abutments (PFM-GA) using two different implant connections, internal butt-joint (IBJ) and internal conical (IC).

Materials and methods: Forty-eight implants (4.3 mm diameter) were divided into four groups (n = 12) to support 2-unit mandibular premolar cantilevers with two different materials (Zr-TiB vs. PFM-GA) and two connection types (IBJ vs. IC). Tested groups were as follows: (1) IBJ/Zr-TiB; (2) IBJ/PFM-GA; (3) IC/Zr-TiB; and (4) IC/PFM-GA. Specimens were thermomechanical aged (1,200,000 cycles, 98 N, 5-55°C) with occlusal axial load on the pontic. Catastrophic and non-catastrophic events were registered, and removal torque values measured before and after aging. Specimens surviving aging were subjected to loading until failure. Survival, total complication rates, torque loss (%), and bending moments were calculated.

Results: From 48 specimens, 38 survived aging. Survival rates significantly varied from 16.7% (IC/PFM-GA) to 100% (IBJ/Zr-TiB; IBJ/PFM-GA; IC/Zr-TiB) (p < .01). Internal conical connection revealed significantly higher torque loss (IC/ZrTiB - 67%) compared to internal butt-joint (IBJ/Zr-TiB - 44%; IBJ/PFM-GA - 46%) (p < .01). Bending moments were higher in internal butt-joint connections than in internal conical (p < .05).

Conclusion and clinical implications: Two-unit posterior implant-supported cantilever FDPs replacing mandibular premolars composed of monolithic zirconia on titanium-base abutments demonstrated higher mechanical stability compared to porcelain-fused-to-metal on castable gold abutments in this in vitro study. The internal conical connection combined with porcelain-fused-to-metal on gold abutments revealed a high number of failures; therefore, their clinical use may be considered cautiously for this indication.

Keywords

  • Aging
  • Bending moment
  • Cantilever
  • Fixed dental prosthesis
  • Implant abutment
  • Implant-abutment connection
  • Zirconia

UNIGE research groups

Citation (ISO format)

RIERA HOMAR, Carme et al. Mechanical stability of posterior implant-supported monolithic zirconia cantilever on titanium-base abutments. An in vitro study. In: Clinical oral implants research, 2024. doi: 10.1111/clr.14251

Main files (1)

Article (Published version)

Identifiers

Journal ISSN

0905-7161
97views
722downloads

Technical informations

Creation

20/06/2024 11:35:59

First validation

21/06/2024 07:19:52

Update

13/10/2025 13:41:30

Status update

21/06/2024 07:19:52

Last indexation

09/11/2026 17:28:42
All rights reserved by Archive ouverte UNIGE and the University of Geneva
OpenAIRE Indexed & Validated