en
Scientific article
English

Influence of sterilization on the retention forces of lithium disilicate and polymer-infiltrated ceramic-network crowns bonded to titanium base abutments : An in-vitro study

Published inClinical oral implants research, vol. 35, no. 4, p. 377-385
Publication date2024-04
First online date2024-01-03
Abstract

Objectives: The aim of this study was to evaluate the effect of sterilization on the retention forces of lithium disilicate (LD) and polymer-infiltrated ceramic network (PICN) crowns bonded to titanium base (Ti-base) abutments.

Materials and methods: Forty LD and 40 PICN crowns were milled and then bonded to 80 Ti-bases with two resin composite cements: Multilink Hybrid Abutment (mh) and Panavia V5 (pv) for a total of 8 groups (n = 10). Half of the specimens (test) underwent an autoclaving protocol (pressure 1.1 bar, 121°C, 20.5 min) and the other half not (control). Restorations were screw-retained to implants, and retention forces (N) were measured with a pull-off testing machine. The surfaces of the Ti-bases and the crowns were inspected for the analysis of the integrity of the marginal bonding interface and failure mode. Student's t-test, chi-square test, and univariate linear regression model were performed to analyze the data (α = 0.05).

Results: The mean pull-off retention forces ranged from 487.7 ± 73.4 N to 742.2 ± 150.3 N. Sterilized groups showed statistically significant overall higher maximum retention forces (p < .05), except for one combination (LD + mh). Sterilization led to an increased presence of marginal gaps and deformities compared to no-sterilization (p < .001), while no statistically significant relationship was found between failure mode and sterilization (p > .05).

Conclusions: Sterilization may have a beneficial effect on the retention forces of LD and PICN crowns bonded to titanium base abutments, although it may negatively influence the integrity of the marginal bonding interface.

eng
Keywords
  • CAD–CAM
  • Bonding
  • Implant dentistry
  • Pull‐off
  • Retention force
  • Sterilization
  • Titanium base abutment
  • Titanium
  • Polymers
  • Dental Porcelain
  • Crowns
  • Materials Testing
  • Zirconium
  • Ceramics
  • Dental Stress Analysis
  • Dental Abutments
  • Computer-Aided Design
Citation (ISO format)
LAFORÌ, Andreina et al. Influence of sterilization on the retention forces of lithium disilicate and polymer-infiltrated ceramic-network crowns bonded to titanium base abutments : An in-vitro study. In: Clinical oral implants research, 2024, vol. 35, n° 4, p. 377–385. doi: 10.1111/clr.14234
Main files (1)
Article (Published version)
accessLevelRestricted
Identifiers
ISSN of the journal0905-7161
14views
0downloads

Technical informations

Creation06/20/2024 11:47:28 AM
First validation06/21/2024 7:08:49 AM
Update time06/21/2024 7:08:49 AM
Status update06/21/2024 7:08:49 AM
Last indexation06/21/2024 7:09:10 AM
All rights reserved by Archive ouverte UNIGE and the University of GenevaunigeBlack