Optimizing vestibular implant electrode positioning using fluoroscopy and intraoperative CT imaging

Publication date2024-01-05
First online date2024-01-05
Abstract

Purpose

Vestibular implant electrode positioning close to the afferent nerve fibers is considered to be key for effective and selective electrical stimulation. However, accurate positioning of vestibular implant electrodes inside the semicircular canal ampullae is challenging due to the inability to visualize the target during the surgical procedure. This study investigates the accuracy of a new surgical protocol with real-time fluoroscopy and intraoperative CT imaging, which facilitates electrode positioning during vestibular implant surgery.

Methods

Single-center case-controlled cohort study with a historic control group at a tertiary referral center. Patients were implanted with a vestibulocochlear implant, using a combination of intraoperative fluoroscopy and cone beam CT imaging. The control group consisted of five patients who were previously implanted with the former implant prototype, without the use of intraoperative imaging. Electrode positioning was analyzed postoperatively with a high-resolution CT scan using 3D slicer software. The result was defined as accurate if the electrode position was within 1.5 mm of the center of the ampulla.

Results

With the new imaging protocol, all electrodes could be positioned within a 1.5 mm range of the center of the ampulla. The accuracy was significantly higher in the study group with intraoperative imaging (21/21 electrodes) compared to the control group without intraoperative imaging (10/15 electrodes), ( p = 0.008).

Conclusion

The combined use of intraoperative fluoroscopy and CT imaging during vestibular implantation can improve the accuracy of electrode positioning. This might lead to better vestibular implant performance.

Keywords
  • Bilateral vestibulopathy
  • Fluoroscopy
  • Imaging
  • Vestibular dysfunction
  • Vestibular implant
Citation (ISO format)
LOOS, Elke et al. Optimizing vestibular implant electrode positioning using fluoroscopy and intraoperative CT imaging. In: European archives of oto-rhino-laryngology, 2024. doi: 10.1007/s00405-023-08428-5
Main files (1)
Article (Published version)
Identifiers
Additional URL for this publicationhttps://link.springer.com/10.1007/s00405-023-08428-5
Journal ISSN0937-4477
93views
109downloads

Technical informations

Creation21/02/2024 09:01:59
First validation29/04/2024 07:29:35
Update21/11/2025 12:48:31
Status update21/11/2025 12:48:31
Last indexation21/11/2025 12:48:42
All rights reserved by Archive ouverte UNIGE and the University of GenevaunigeBlack