Effect of splinting scan bodies on the trueness of complete-arch digital implant scans with 5 different intraoral scanners

Publication date2023-08-01
First online date2023-08-01
Abstract

Statement of problem: The absence of fixed reference points can affect the trueness of complete-arch intraoral digital implant scans. The effect of splinting intraoral scan bodies (ISBs) or the inclusion of artificial landmarks (AL) on the trueness of complete-arch digital implant scans is still unclear.

Purpose: The purpose of this study was to analyze the effect of splinting ISBs or the inclusion of AL on the trueness of complete-arch digital implant scans with 5 intraoral scanners (IOSs).

Material and methods: Six tissue-level dental implants (Straumann Tissue Level) were placed in an edentulous patient, and the correspondent definitive cast was digitized with a desktop scanner (IScan4D LS3i) to obtain the reference digital cast. Digital scans (n=10) were performed with 5 IOSs: TRIOS 4, Virtuo Vivo, Medit i700, iTero Element 5D, and Cerec Primescan. Three different scanning techniques were evaluated: conventional (cIOSs), splinted (sIOSs), and AL (AL-IOSs). The scan data obtained were imported into a metrology software program and superimposed to the reference digital cast by using a best-fit algorithm. The overall deviations of the positions of the ISBs were evaluated by using the root-mean-square (RMS) error (α=.05).

Results: The mean ±standard deviation trueness values for the cIOSs, sIOSs, and AL-IOSs groups were 48 ±8 µm, 53 ±7 µm, and 49 ±11 µm, respectively, with no statistically significant differences (P=.06). Significant differences were found for the IOSs used with each technique (P<.001). Primescan (27 ±4 µm cIOSs; 28 ±3 µm sIOSs; 31 ±3 µm AL-IOSs) showed significantly higher trueness than iTero 5D (47 ±5 µm cIOSs; 47 ±4 µm sIOSs; 50 ±6 µm AL-IOSs) (P=.002) and TRIOS 4 (93 ±18 µm cIOSs; 76 ±18 µm sIOSs; 107 ±13 µm AL-IOSs) (P=.001) for all techniques. In addition, no significant differences were found between the techniques by using iTero 5D or Primescan (P=.348 and P=.059, respectively).

Conclusions: The cIOSs, sIOSs, and AL-IOSs techniques showed similar trueness. The IOS used influenced the trueness of complete-arch digital implant scans.

Citation (ISO format)
AZEVEDO, Luis Pedro et al. Effect of splinting scan bodies on the trueness of complete-arch digital implant scans with 5 different intraoral scanners. In: The journal of prosthetic dentistry, 2023. doi: 10.1016/j.prosdent.2023.06.015
Main files (1)
Article (Published version)
Identifiers
Journal ISSN0022-3913
89views
177downloads

Technical informations

Creation09/21/2023 12:26:36 PM
First validation09/25/2023 1:38:31 PM
Update time09/25/2023 1:38:31 PM
Status update09/25/2023 1:38:31 PM
Last indexation11/01/2024 6:08:25 AM
All rights reserved by Archive ouverte UNIGE and the University of GenevaunigeBlack