Scientific article
English

Optimization of cardiac cine in the rat on a clinical 1.5-T MR system

Published inMagma, vol. 19, no. 3, p. 144-151
First online date2006-07-25
Abstract

Object: the overall goal was to study cardiovascular function in small animals using a clinical 1.5-T MR scanner optimizing a fast gradient-echo cine sequence to obtain high spatial and temporal resolution.

Materials and methods: normal rat hearts (n = 9) were imaged using a 1.5-T MR scanner with a spiral fast gradient-echo (fast field echo for Philips scanners) sequence, three Cartesian fast gradient-echo (turbo field echo for Philips scanners) sequences with different in-plane resolution, and with and without flow compensation and half-Fourier acquisition. The hearts of four rats were then excised and left-ventricle mass was weighed. Inter- and intra-observer variability analysis was performed for magnetic resonance imaging (MRI) measurements.

Results: half-Fourier acquisition with flow compensation gave the best sequence in terms of image quality, spatial as well as temporal resolution, and suppression of flow artifact. Ejection fraction was 71 ± 4% with less than 5% inter- and intra-observer variability. A good correlation was found between MRI-calculated left-ventricular mass and wet weight.

Conclusions: using optimized sequences on a clinical 1.5-T MR scanner can provide accurate quantification of cardiac function in small animals and can promote cardiovascular research on small animals at 1.5-T

Keywords
  • Algorithms
  • Animals
  • Heart Ventricles / anatomy & histology
  • Image Enhancement / methods
  • Image Interpretation, Computer-Assisted
  • Magnetic Resonance Imaging / instrumentation
  • Magnetic Resonance Imaging / methods
  • Male
  • Organ Size
  • Rats
  • Rats, Sprague-Dawley
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Stroke Volume / physiology
  • Ventricular Function
  • Ventricular Function, Left / physiology
  • Video Recording / methods
Citation (ISO format)
MONTET-ABOU, Karin et al. Optimization of cardiac cine in the rat on a clinical 1.5-T MR system. In: Magma, 2006, vol. 19, n° 3, p. 144–151. doi: 10.1007/s10334-006-0037-z
Main files (1)
Article (Published version)
accessLevelRestricted
Identifiers
Additional URL for this publicationhttp://link.springer.com/10.1007/s10334-006-0037-z
Journal ISSN0968-5243
87views
0downloads

Technical informations

Creation29/11/2022 10:21:45
First validation05/05/2023 07:32:17
Update05/05/2023 07:32:17
Status update05/05/2023 07:32:17
Last indexation01/11/2024 04:55:30
All rights reserved by Archive ouverte UNIGE and the University of GenevaunigeBlack