en
Scientific article
Open access
English

In vitro and in vivo accuracy of full-arch digital implant impressions

Published inClinical oral implants research, vol. 32, no. 12, p. 1444-1454
Publication date2021-12
First online date2021-09-30
Abstract

Objectives: The main objective of the study was to compare the accuracy of full-arch digital implant impressions for fixed dental prosthesis under in vitro and in vivo conditions.

Materials and methods: Eight patients (five women and three men) with at least one edentulous arch and with 4-6 osseointegrated implants participated in this study. For each edentulous arch (n = 10), experimental screw-retained titanium bar with attached four scan bodies was fabricated. The bar containing four scan bodies was screw-retained intraorally on implants and scanned with Trios 3 intraoral scanner eight times (IOS group, in vivo). Then, the bar was attached to the master cast and scanned eight times again with the same intraoral scanner (MIOS group, in vitro). Finally, the bar with scan bodies was scanned 8 times with a laboratory scanner (reference). Precision and trueness were calculated for 3 distances and 3 angles between the scan bodies (1-2, 1-3, and 1-4) in IOS and MIOS groups.

Results: Precision and trueness for the largest distance (1-4) were found to be 44 ± 18 µm and 32 ± 19 µm for the IOS group and 31 ± 16 µm and 30 ± 14 µm for MIOS group, respectively. Precision and trueness for the angle between the most distant scan bodies (1-4) were 0.22 ± 0.14° and 0.18 ± 0.10° for the IOS group and 0.16 ± 0.11° and 0.07 ± 0.05° for MIOS group, respectively.

Conclusions: Intraoral conditions moderately affected the precision and trueness of Trios 3 (3Shape) intraoral scanner. Results of in vitro accuracy studies cannot be directly transferred to the clinical field.

eng
Keywords
  • Clinical assessment
  • Diagnosis
  • Patient centered outcomes
  • Prosthodontics
  • Computer-Aided Design
  • Dental Implants
  • Dental Impression Technique
  • Humans
  • Imaging, Three-Dimensional
  • Models, Dental
Citation (ISO format)
RUTKUNAS, Vygandas et al. In vitro and in vivo accuracy of full-arch digital implant impressions. In: Clinical oral implants research, 2021, vol. 32, n° 12, p. 1444–1454. doi: 10.1111/clr.13844
Main files (1)
Article (Published version)
Identifiers
ISSN of the journal0905-7161
102views
65downloads

Technical informations

Creation11/22/2022 8:33:00 AM
First validation11/22/2022 8:33:00 AM
Update time03/16/2023 9:03:47 AM
Status update03/16/2023 9:03:46 AM
Last indexation05/06/2024 2:38:08 PM
All rights reserved by Archive ouverte UNIGE and the University of GenevaunigeBlack