Scientific article
English

Phenytoin speciation with potentiometric and chronopotentiometric ion-selective membrane electrodes

Published in

Biosensors and Bioelectronics, vol. 79, p. 114-120

Publication date

2016

Abstract

We report on an electrochemical protocol based on perm-selective membranes to provide valuable information about the speciation of ionizable drugs, with phenytoin as a model example. Membranes containing varying amounts of tetradodecylammonium chloride (TDDA) were read out at zero current (potentiometry) and with applied current techniques (chronopotentiometry). Potentiometry allows one to assess the ionized form of phenytoin (pKa~8.2) that corresponds to a negatively monocharged ion. A careful optimization of the membrane components resulted in a lower limit of detection (~1.6 µM) than previous reports. Once the pH (from 9 to 10) or the concentration of albumin is varied in the sample (from 0 to 30 g L−1), the potentiometric signal changes abruptly as a result of reducing/increasing the ionized concentration of phenytoin. Therefore, potentiometry as a single technique is by itself not sufficient to obtain information about the concentration and speciation of the drug in the system. For this reason, a tandem configuration with chronopotentiometry as additional readout principle was used to determine the total and ionized concentration of phenytoin. In samples containing excess albumin the rate-limiting step for the chronopotentiometry readout appears to be the diffusion of ionized phenytoin preceded by comparatively rapid deprotonation and decomplexation reactions. This protocol was applied to measure phenytoin in pharmaceutical tables (100 mg per tablet). This tandem approach can likely be extended to more ionizable drugs and may eventually be utilized in view of pharmacological monitoring of drugs during the delivery process.

UNIGE research groups

Citation (ISO format)

JANSOD, Sutida et al. Phenytoin speciation with potentiometric and chronopotentiometric ion-selective membrane electrodes. In: Biosensors and Bioelectronics, 2016, vol. 79, p. 114–120. doi: 10.1016/j.bios.2015.12.011

Main files (1)

Article (Published version)

Identifiers

Journal ISSN

0956-5663
540views
1downloads

Technical informations

Creation

19/10/2017 15:10:00

First validation

19/10/2017 15:10:00

Update

15/03/2023 02:09:57

Status update

15/03/2023 02:09:57

Last indexation

31/10/2024 08:11:02
All rights reserved by Archive ouverte UNIGE and the University of Geneva
OpenAIRE Indexed & Validated