Time-of-flight (TOF) PET/MR imaging is an emerging imaging technology with great capabilities offered by TOF to improve image quality and lesion detectability. We assessed, for the first time, the impact of TOF image reconstruction on PET quantification errors induced by MR imaging-based attenuation correction (MRAC) using simulation and clinical PET/CT studies.

Keywords
  • Aged
  • Algorithms
  • Artifacts
  • Computer Simulation
  • Female
  • Fluorodeoxyglucose F18/diagnostic use
  • Humans
  • Image Processing, Computer-Assisted/methods
  • Magnetic Resonance Imaging/methods
  • Male
  • Metals
  • Middle Aged
  • Multimodal Imaging
  • Positron-Emission Tomography/methods
  • Reproducibility of Results
  • Respiration
  • Tomography, X-Ray Computed/methods
  • Whole Body Imaging/methods
Citation (ISO format)
MEHRANIAN, Abolfazl, ZAIDI, Habib. Impact of time-of-flight PET on quantification errors in MR imaging-based attenuation correction. In: The Journal of nuclear medicine, 2015, vol. 56, n° 4, p. 635–641. doi: 10.2967/jnumed.114.148817
Main files (1)
Article (Published version)
accessLevelRestricted
Identifiers
ISSN of the journal0161-5505
487views
1downloads

Technical informations

Creation12/03/2015 2:03:00 PM
First validation12/03/2015 2:03:00 PM
Update time03/14/2023 11:59:13 PM
Status update03/14/2023 11:59:13 PM
Last indexation01/16/2024 7:51:20 PM
All rights reserved by Archive ouverte UNIGE and the University of GenevaunigeBlack