Corneal Biomechanical Properties at Different Corneal Cross-Linking (CXL) Irradiances

Published in

Investigative ophthalmology & visual science, vol. 55, no. 5, p. 2881-2884

Publication date

2014

Abstract

Purpose: New corneal cross-linking (CXL) devices are capable of using higher UV-A light irradiances than used in original CXL protocols. The Bunsen-Roscoe law states that a photochemical reaction should stay constant if the delivered total energy is kept constant; however, little clinical data are available to support this hypothesis.

Methods: We investigated the biomechanical properties of four groups (n = 50 each) of porcine corneas. Three groups were exposed to riboflavin 0.1 % and UV-A irradiation of equal total energy (3 mW/cm(2) for 30 minutes, 9 mW/cm(2) for 10 minutes, and 18 mW/cm(2) for 5 minutes). Controls were exposed to riboflavin 0.1% without irradiation. Young's modulus of 5-mm wide corneal strips was used as an indicator of corneal stiffness.

Results: We observed a decreased stiffening effect with increasing UV-A intensity. Young's modulus at 10% strain showed significant differences between 3 mW/cm(2) and 9 mW/cm(2) (P = 0.002), 3 mW/cm(2) and 18 mW/cm(2) (P = 0.0002), 3 mW/cm(2) and the control group (P < 0.0001), and 9 mW/cm(2) and the control group (P = 0.015). There was no difference between 18 mW/cm(2) and the control group (P = 0.064) and between 9 mW/cm(2) and 18 mW/cm(2) (P = 0.503).

Conclusions: The biomechanical effect of CXL decreased significantly when using high irradiance/short irradiation time settings. Intrastromal oxygen diffusion capacity and increased oxygen consumption associated with higher irradiances may be a limiting factor leading to reduced treatment efficiency. Our results regarding the efficiency of high-irradiance collagen cross-linking (CXL) raise concerns about the clinical efficiency of the new high-irradiance CXL devices already used in clinical practice without proper validation.

Keywords

  • Biomechanics
  • Corneal cross-linking
  • Efficiency
  • High irradiance
  • Oxygen

Citation (ISO format)

HAMMER, Arthur et al. Corneal Biomechanical Properties at Different Corneal Cross-Linking (CXL) Irradiances. In: Investigative ophthalmology & visual science, 2014, vol. 55, n° 5, p. 2881–2884. doi: 10.1167/iovs.13-13748

Main files (1)

Article (Published version)

Identifiers

Journal ISSN

0146-0404
581views
886downloads

Technical informations

Creation

14/10/2015 14:18:00

First validation

14/10/2015 14:18:00

Update

11/10/2023 09:28:29

Status update

11/10/2023 09:28:29

Last indexation

31/10/2024 01:58:27
All rights reserved by Archive ouverte UNIGE and the University of Geneva
OpenAIRE Indexed & Validated