UNIGE document Scientific Article
previous document  unige:45632  next document
add to browser collection
Title

New Generation Neonatal High Frequency Ventilators: Effect of Oscillatory Frequency and Working Principles on Performance

Authors
Published in Respiratory Care. 2015, vol. 60, no. 3, p. 363-370
Abstract Several new generation neonatal ventilators that incorporate conventional as well as high frequency ventilation (HFOV) have appeared on the market. Most of them offer the possibility to use HFOV in a volume-targeted mode, despite absence of any preclinical data. With a bench test, we evaluated the performances of 4 new neonatal HFOV devices and compared them to the SensorMedics HFOV device.
Identifiers
PMID: 25406342
Full text
Article (Published version) (1.4 MB) - document accessible for UNIGE members only Limited access to UNIGE
Structures
Research group Maladies pulmonaires aigues et pathologies cardiaques congénitales (697)
Citation
(ISO format)
GRAZIOLI, Serge, KARAM, Oliver Laurent, RIMENSBERGER, Peter. New Generation Neonatal High Frequency Ventilators: Effect of Oscillatory Frequency and Working Principles on Performance. In: Respiratory Care, 2015, vol. 60, n° 3, p. 363-370. https://archive-ouverte.unige.ch/unige:45632