Abstract

We present a novel optical aerosol particle detector based on Xe flash lamp excitation and spectrally resolved fluorescence acquisition. We demonstrate its performances on three natural pollens acquiring in real-time scattering intensity at two wavelengths, sub-microsecond time-resolved scattering traces of the particles' passage in the focus, and UV-excited fluorescence spectra. We show that the device gives access to a rather specific detection of the bioaerosol particles.

Keywords

  • Aerosols
  • Fluorescence spectra
  • Pollen
  • Particle scattering
  • Laminar flows

Citation (ISO format)

KISELEV, Denis, BONACINA, Luigi, WOLF, Jean-Pierre. A flash-lamp based device for fluorescence detection and identification of individual pollen grains. In: Review of scientific instruments, 2013, vol. 84, n° 3, p. 033302. doi: 10.1063/1.4793792