en
Scientific article
Open access
English

Ten-year results with the Morscher press-fit cup: an uncemented, non-modular, porous-coated cup inserted without screws

Published inInternational orthopaedics, vol. 35, no. 7, p. 957-963
Publication date2011
Abstract

Total hip arthroplasty (THA) with well designed cementless acetabular implants has shown excellent results. The purpose of this study was to assess our clinical and radiological outcomes using an uncemented cup. We conducted a prospective cohort study including all consecutive primary THAs performed with the Morscher press-fit cup, an uncemented non-modular acetabular component, between March 1996 and December 1998. Patients were evaluated at ten years with clinical and radiological follow-up, patient satisfaction and questionnaire assessment using the Harris hip score (HHS), Merle d'Aubigné and Postel score, the UCLA score, the 12-item short-form health survey (SF-12) and a visual analog scale. Five hundred sixty-one THAs were performed in 518 patients. At 120 months (± 7.3 months), 303 patients with 335 THAs were still available for follow-up. None of the patients had required cup revision for aseptic loosening. At ten years, the cup survivorship was 98.8% (95% CI 97.4-99.5) with cup revision for any cause as an endpoint. No radiolucencies were seen around the cups, but osteolytic defects involved 21 stems (8.3%). Mean total linear polyethylene wear was 0.9 mm. The Morscher acetabular replacement cup provides excellent results at ten years. There were no revisions for aseptic loosening of the cup, and no osteolytic defects were found around the cup. Patient satisfaction was high and the clinical results were very good.

Keywords
  • Acetabulum/radiography/surgery
  • Adult
  • Aged
  • Aged, 80 and over
  • Arthroplasty, Replacement, Hip/adverse effects/instrumentation/methods
  • Cementation
  • Female
  • Follow-Up Studies
  • Health Status
  • Hip Joint/physiopathology/radiography/surgery
  • Hip Prosthesis
  • Humans
  • Kaplan-Meier Estimate
  • Male
  • Middle Aged
  • Osteolysis/etiology
  • Patient Satisfaction
  • Postoperative Complications
  • Prospective Studies
  • Prosthesis Design
  • Treatment Outcome
  • Young Adult
Citation (ISO format)
GARAVAGLIA, Guido et al. Ten-year results with the Morscher press-fit cup: an uncemented, non-modular, porous-coated cup inserted without screws. In: International orthopaedics, 2011, vol. 35, n° 7, p. 957–963. doi: 10.1007/s00264-010-1059-8
Main files (1)
Article (Published version)
accessLevelPublic
Identifiers
ISSN of the journal0341-2695
630views
1193downloads

Technical informations

Creation11/05/2012 11:59:00 AM
First validation11/05/2012 11:59:00 AM
Update time03/14/2023 5:48:38 PM
Status update03/14/2023 5:48:38 PM
Last indexation10/18/2023 3:06:41 PM
All rights reserved by Archive ouverte UNIGE and the University of GenevaunigeBlack