Navigator-based motion compensation for liver BOLD measurement with five-echo SAGE EPI and breath-hold task

Published inNMR in biomedicine, vol. 37, no. 10, e5173
Publication date2024-10
First online date2024-05-24
Abstract

Purpose: The purpose of this work is to apply multi-echo spin- and gradient-echo (SAGE) echo-planar imaging (EPI) combined with a navigator-based (NAV) prospective motion compensation method for a quantitative liver blood oxygen level dependent (BOLD) measurement with a breath-hold (BH) task.

Methods: A five-echo SAGE sequence was developed to quantitatively measure T2 and T2 * to depict function with sufficient signal-to-noise ratio, spatial resolution and sensitivity to BOLD changes induced by the BH task. To account for respiratory motion, a navigator was employed in the form of a single gradient-echo projection readout, located at the diaphragm along the inferior-superior direction. Prior to each transverse imaging slice of the spin-echo EPI-based readouts, navigator acquisition and fat suppression were incorporated. Motion data was obtained from the navigator and transmitted back to the sequence, allowing real-time adjustments to slice positioning. Six healthy volunteers and three patients with liver carcinoma were included in this study. Quantitative T2 and T2 * were calculated at each time point of the BH task. Parameters of t value from first-level analysis using a general linear model and hepatovascular reactivity (HVR) of Echo1, T2 and T2 * were calculated.

Results: The motion caused by respiratory activity was successfully compensated using the navigator signal. The average changes of T2 and T2 * during breath-hold were about 1% and 0.7%, respectively. With the help of NAV prospective motion compensation whole liver t values could be obtained without motion artifacts. The quantified liver T2 (34.7 ± 0.7 ms) and T2 * (29 ± 1.2 ms) values agreed with values from literature. In healthy volunteers, the distribution of statistical t value and HVR was homogeneous throughout the whole liver. In patients with liver carcinoma, the distribution of t value and HVR was inhomogeneous due to metastases or therapy.

Conclusions: This study demonstrates the feasibility of using a NAV prospective motion compensation technique in conjunction with five-echo SAGE EPI for the quantitative measurement of liver BOLD with a BH task.

Keywords
  • Hepatovascular reactivity
  • Navigator‐based slice tracking
  • Prospective motion compensation
  • Quantitative liver BOLD
  • Humans
  • Echo-Planar Imaging
  • Breath Holding
  • Liver / diagnostic imaging
  • Male
  • Adult
  • Female
  • Middle Aged
  • Motion
  • Oxygen / blood
  • Liver Neoplasms / diagnostic imaging
  • Aged
Funding
  • Deutsche Forschungsgemeinschaft [507778062]
  • German Federal Ministry of Education and Research [82DZL004A1]
Citation (ISO format)
ZHANG, Ke et al. Navigator-based motion compensation for liver BOLD measurement with five-echo SAGE EPI and breath-hold task. In: NMR in biomedicine, 2024, vol. 37, n° 10, p. e5173. doi: 10.1002/nbm.5173
Main files (1)
Article (Published version)
Secondary files (1)
Supplemental data
accessLevelPublic
Identifiers
Journal ISSN0952-3480
2views
9downloads

Technical informations

Creation26/05/2026 13:58:59
First validation19/06/2026 06:52:00
Update19/06/2026 06:52:00
Status update19/06/2026 06:52:00
Last indexation19/06/2026 06:52:01
All rights reserved by Archive ouverte UNIGE and the University of GenevaunigeBlack