Scientific article
OA Policy
English

Performance Evaluation of the Time-of-Flight Biograph Vision.X : A Fast Coincidence Time Resolution PET/CT Scanner

Published inIEEE transactions on radiation and plasma medical sciences, vol. 10, no. 6, p. 951-958
Publication date2026-07
First online date2025-12-08
Abstract

This work evaluated the performance characteristics of the Biograph Vision.X PET/CT system according to the NEMA NU 2-2018 standards. This novel system is based on the Biograph Vision 600 with an axial field-of-view of 26.1 cm, consisting of 8 rings of 38 detector blocks using the same detector modules combining 5 × 5 array of 3.2 × 3.2 × 20 mm lutetium-oxyorthosilicate crystals with 16 × 16 mm array of silicon photomultiplier–based photodetectors. The Biograph Vision.X scanner is a newly released time-of-flight whole-body hybrid PET/CT imaging system featuring exclusive proprietary high-performance detector electronic assembly (DEA) technology with optimized digital electronics and digital processing along the entire imaging chain to achieve fast coincidence time resolution PET. In addition, examples of 3D Hoffman phantom and clinical 18F-FDG brain and whole-body studies were compared with scans performed sequentially after the Biograph Vision 600 PET/CT scanner. The Biograph Vision.X had a radial, tangential and axial spatial resolution in terms of full-width at half-maximum of 3.4, 3.5 and 3.5 mm, respectively, at a 1 cm offset from the center of the field-of-view. The sensitivity was 19.2 cps/kBq at the center and 10 cm off-center, while the peak noise-equivalent count-rate (NECR) was 291 kcps at 27.2 kBq/mL. The TOF resolution varied from 178.7 to 182.0 ps as the count rate increased up to the peak NECR. Generally, the image contrast observed on the NEMA image quality phantom ranged from 74.1% to 91.8%. The Biograph Vision.X is a competitive newly introduced device outperforming its predecessor, the Biograph Vision 600, by reaching the fastest TOF performance in the industry, thus achieving 7.5 TOF gain improvement compared to 6.2 on the Vision 600.

Keywords
  • PET/CT
  • Time-of-flight
  • Performance evaluation
  • NEMA
  • Image quality
Citation (ISO format)
ZAIDI, Habib et al. Performance Evaluation of the Time-of-Flight Biograph Vision.X : A Fast Coincidence Time Resolution PET/CT Scanner. In: IEEE transactions on radiation and plasma medical sciences, 2026, vol. 10, n° 6, p. 951–958. doi: 10.1109/TRPMS.2025.3640401
Main files (1)
Article (Published version)
Identifiers
Additional URL for this publicationhttps://ieeexplore.ieee.org/document/11284795/
Journal ISSN2469-7303
13views
195downloads

Technical informations

Creation05/02/2026 08:50:50
First validation13/02/2026 15:36:57
Update08/07/2026 07:32:07
Status update08/07/2026 07:32:07
Last indexation08/07/2026 07:32:09
All rights reserved by Archive ouverte UNIGE and the University of GenevaunigeBlack