Image reconstruction and elongation artifact reduction for a dual‐panel dedicated prostate PET scanner

Published inMedical physics, vol. 53, no. 2, e70298
Publication date2026-02
First online date2026-01-27
Abstract

Background : The development of PET scanners dedicated to high temporal and spatial resolution organ‐specific imaging is an active research area, motivated by the need for cost reduction, improved lesion detectability and quantification in specific clinical scenarios, as well as by ongoing hardware and software innovations.

Purpose : This study investigates and compares various image reconstruction strategies for a dual‐panel prostate‐dedicated PET scanner (ProVision), which features four‐layered dual‐readout time‐of‐flight depth‐of‐interaction detectors and a 22‐position acquisition protocol to improve angular coverage.

Materials and methods : A list‐mode MLEM algorithm with multi‐ray modeling was developed and optimized using a scaled NEMA image quality phantom to determine optimal number of rays and iterations. These parameters were then used to reconstruct data from both simulation and experimental acquisitions, including an anthropomorphic pelvis phantom, named Adam‐PETer. Four reconstruction approaches were evaluated: classical MLEM; MLEM with embedded shift‐variant point spread function (PSF) modeling; a hybrid list‐mode reconstruction; and a Swin‐UNETR‐based deep learning model applied as a post‐reconstruction enhancement to MLEM images. Performance was assessed using contrast recovery coefficient (CRC), contrast‐to‐noise ratio (CNR), and contrast‐to‐noise consistency (CNC), on both a scaled NEMA phantom and an experimental anthropomorphic phantom.

Results : In the scaled NEMA phantom simulation, the Swin‐based method yielded the highest CNR and CNC, especially for the smallest spheres, thereby outperforming both standard MLEM and the hybrid algorithm. In the Adam‐PETer experimental prostate phantom, the CNR was 10.43 for MLEM, 14.48 for Hybrid, and 13.85 for Swin for the larger lesion (10 mm). The CNR values were 2.28, 3.03, and 4.35, respectively, for the smaller lesion (8 mm). CNC values also varied across methods, with Swin achieving the best result for the smaller lesion. These findings indicate that model‐based and learned methods offer complementary strengths depending on lesion contrast and size.

Conclusion : The PET scanner‐adapted reconstruction combined with deep learning refinement improves image quality in dedicated, limited‐angle PET systems.

Keywords
  • Artifacts
  • Humans
  • Image Processing, Computer-Assisted / methods
  • Male
  • Phantoms, Imaging
  • Positron-Emission Tomography / instrumentation
  • Prostate / diagnostic imaging
Citation (ISO format)
SABERI MANESH, Abdollah et al. Image reconstruction and elongation artifact reduction for a dual‐panel dedicated prostate PET scanner. In: Medical physics, 2026, vol. 53, n° 2, p. e70298. doi: 10.1002/mp.70298
Main files (1)
Article (Published version)
Secondary files (1)
Supplemental data
accessLevelPublic
Identifiers
Journal ISSN0094-2405
15views
48downloads

Technical informations

Creation27/01/2026 14:28:43
First validation04/02/2026 08:47:41
Update04/02/2026 08:47:41
Status update04/02/2026 08:47:41
Last indexation05/02/2026 09:18:26
All rights reserved by Archive ouverte UNIGE and the University of GenevaunigeBlack