Valved stents for transapical pulmonary valve replacement

Published inJournal of thoracic and cardiovascular surgery, vol. 137, no. 4, p. 914-918
Publication date2009-04
First online date2008-12-19
Abstract

Objectives: Pulmonary valve insufficiency remains a leading cause for reoperations in congenital cardiac surgery. The current percutaneous approach is limited by the size of the access vessel and variable right ventricular outflow tract morphology. This study assesses the feasibility of transapical pulmonary valve replacement based on a new valved stent construction concept.

Methods: A new valved stent design was implanted off-pump under continuous intracardiac echocardiographic and fluoroscopic guidance into the native right ventricular outflow tract in 8 pigs (48.5 +/- 6.0 kg) through the right ventricular apex, and device function was studied by using invasive and noninvasive measures.

Results: Procedural success was 100% at the first attempt. Procedural time was 75 +/- 15 minutes. All devices were delivered at the target site with good acute valve function. No valved stents dislodged. No animal had significant regurgitation or paravalvular leaking on intracardiac echocardiographic analysis. All animals had a competent tricuspid valve and no signs of right ventricular dysfunction. The planimetric valve orifice was 2.85 +/- 0.32 cm(2). No damage to the pulmonary artery or structural defect of the valved stents was found at necropsy.

Conclusions: This study confirms the feasibility of direct access valve replacement through the transapical procedure for replacement of the pulmonary valve, as well as validity of the new valved stent design concept. The transapical procedure is targeting a broader patient pool, including the very young and the adult patient. The device design might not be restricted to failing conduits only and could allow for implantation in a larger patient population, including those with native right ventricular outflow tract configurations.

Keywords
  • Feasibility Studies
  • Heart Valve Prosthesis
  • Heart Valve Prosthesis Implantation / methods
  • Humans
  • Prosthesis Design
  • Pulmonary Valve / surgery
  • Pulmonary Valve Insufficiency / surgery
  • Stents
Affiliation entities Not a UNIGE publication
Citation (ISO format)
HUBER, Christoph et al. Valved stents for transapical pulmonary valve replacement. In: Journal of thoracic and cardiovascular surgery, 2009, vol. 137, n° 4, p. 914–918. doi: 10.1016/j.jtcvs.2008.09.024
Main files (1)
Article (Published version)
Identifiers
Journal ISSN0022-5223
9views
10downloads

Technical informations

Creation26/12/2025 15:22:35
First validation22/01/2026 14:55:49
Update24/03/2026 09:56:20
Status update24/03/2026 09:56:20
Last indexation24/03/2026 09:59:03
All rights reserved by Archive ouverte UNIGE and the University of GenevaunigeBlack