Scientific article
OA Policy
English

Percutaneous management of vascular complications in patients undergoing transcatheter aortic valve implantation

Published inJACC. Cardiovascular interventions, vol. 5, no. 5, p. 515-524
Publication date2012-05
Abstract

Objectives: This study sought to investigate the feasibility and safety of percutaneous management of vascular complications after transcatheter aortic valve implantation (TAVI).

Background: Vascular complications after TAVI are frequent and outcomes after percutaneous management of these adverse events not well established.

Methods: Between August 2007 and July 2010, 149 patients underwent transfemoral TAVI using a percutaneous approach. We compared outcomes of patients undergoing percutaneous management of vascular complications with patients free from vascular complications and performed duplex ultrasonography, fluoroscopy, and multislice computed tomography during follow-up.

Results: A total of 27 patients (18%) experienced vascular complications consisting of incomplete arteriotomy closure (n = 19, 70%), dissection (n = 3, 11%), arterial perforation (n = 3, 11%), arterial occlusion (n = 1, 4%), and pseudoaneurysm (n = 1, 4%). Percutaneous stent graft implantation was successful in 21 of 23 (91%) patients, whereas 2 patients were treated by manual compression, 2 patients underwent urgent surgery, and 2 patients required delayed surgery. Rates of major adverse cardiac events at 30 days were similar among patients undergoing percutaneous management of vascular complications and those without vascular complications (9% vs. 8%, p = 1.00). After a median follow-up of 10.9 months, imaging showed no evidence of hemodynamically significant stenosis (mean peak velocity ratio: 1.2 ± 0.4). Stent fractures were observed in 4 stents (22%, type I [6%], type II [16%]) and were clinically silent in all cases.

Conclusions: Vascular complications after TAVI can be treated percutaneously as a bailout procedure with a high rate of technical success, and clinical outcomes are comparable to patients without vascular complications. Stent patency is high during follow-up, although stent fractures require careful scrutiny.

Keywords
  • Aged
  • Aged, 80 and over
  • Aortic Valve Stenosis / therapy
  • Blood Vessel Prosthesis
  • Blood Vessel Prosthesis Implantation / adverse effects
  • Blood Vessel Prosthesis Implantation / instrumentation
  • Cardiac Catheterization / adverse effects
  • Chi-Square Distribution
  • Endovascular Procedures / adverse effects
  • Endovascular Procedures / instrumentation
  • Feasibility Studies
  • Female
  • Fluoroscopy
  • Heart Valve Prosthesis Implantation / adverse effects
  • Humans
  • Logistic Models
  • Male
  • Multidetector Computed Tomography
  • Predictive Value of Tests
  • Prosthesis Failure
  • Registries
  • Risk Assessment
  • Risk Factors
  • Stents
  • Switzerland
  • Time Factors
  • Treatment Outcome
  • Ultrasonography, Doppler, Duplex
  • Vascular Diseases / diagnosis
  • Vascular Diseases / etiology
  • Vascular Diseases / surgery
  • Vascular Diseases / therapy
Affiliation entities Not a UNIGE publication
Citation (ISO format)
STORTECKY, Stefan et al. Percutaneous management of vascular complications in patients undergoing transcatheter aortic valve implantation. In: JACC. Cardiovascular interventions, 2012, vol. 5, n° 5, p. 515–524. doi: 10.1016/j.jcin.2012.01.021
Main files (1)
Article (Published version)
Identifiers
Journal ISSN1876-7605
7views
16downloads

Technical informations

Creation26/12/2025 15:22:04
First validation22/01/2026 10:36:54
Update22/01/2026 10:36:54
Status update22/01/2026 10:36:54
Last indexation22/01/2026 10:36:55
All rights reserved by Archive ouverte UNIGE and the University of GenevaunigeBlack