Transapical access closure : the TA PLUG device

Published inInteractive cardiovascular and thoracic surgery, vol. 17, no. 5, p. 806-809; discussion 809-810
Publication date2013-11
First online date2013-07-09
Abstract

Objectives: Percutaneous closure of the transapical (TA) access site for large-calibre devices is an unsolved issue. We report the first experimental data on the TA PLUG device for true-percutaneous closure following large apical access for transcatheter aortic valve implantation.

Methods: The TA PLUG, a self-sealing full-core closure device, was implanted in an acute animal study in six pigs (60.2 ± 0.7 kg). All the pigs received 100 IU/kg of heparin. The targeted activated clotting time was left to normalize spontaneously. After accessing the left ventricular apex with a 39 French introducer, the closure plug device was delivered with a 33 French over-the-wire system under fluoroscopic guidance into the apex. Time to full haemostasis as well as rate of bleeding was recorded. Self-anchoring properties were assessed by haemodynamic push stress under adrenalin challenge. An additional feasibility study was conducted in four pigs (58.4 ± 1.1 kg) with full surgical exposure of the apex, and assessed device anchoring by pull-force measurements with 0.5 Newton (N) increments. All the animals were electively sacrified. Post-mortem analysis of the heart was performed and the renal embolic index assessed.

Results: Of six apical closure devices, five were correctly inserted and fully deployed at the first attempt. One became blocked in the delivery system and was placed successfully at the second attempt. In all the animals, complete haemostasis was immediate and no leak was recorded during the 5-h observation period. Neither leak nor any device dislodgement was observed under haemodynamic push stress with repeated left ventricular peak pressure of up to 220 mmHg. In the feasibility study assessing pull-stressing, device migration occurred at a force of 3.3 ± 0.5 N corresponding to 247.5 mmHg. Post-mortem analyses confirmed full expansion of all devices at the intended target. No macroscopic damage was identified at the surrounding myocardium. The renal embolic index was zero.

Conclusions: True-percutaneous left ventricular apex closure following large access is feasible with the self-sealing TA PLUG. The device allows for immediate haemostasis and a reliable anchoring in the acute animal setting. This is the first report of a true-percutaneous closure for large-calibre transcatheter aortic valve implantation access.

Keywords
  • Apex
  • Closure device
  • Percutaneous
  • Sutureless
  • Transapical
  • Transcatheter
  • Transcatheter aortic valve implantation
  • Animals
  • Cardiac Catheterization / adverse effects
  • Cardiac Catheterization / instrumentation
  • Cardiac Catheters
  • Equipment Design
  • Feasibility Studies
  • Foreign-Body Migration / etiology
  • Heart Valve Prosthesis Implantation / adverse effects
  • Heart Valve Prosthesis Implantation / instrumentation
  • Heart Valve Prosthesis Implantation / methods
  • Hemodynamics
  • Hemorrhage / etiology
  • Hemorrhage / physiopathology
  • Hemorrhage / prevention & control
  • Hemostatic Techniques / instrumentation
  • Materials Testing
  • Models, Animal
  • Punctures
  • Stress, Mechanical
  • Swine
  • Time Factors
Affiliation entities Not a UNIGE publication
Citation (ISO format)
BRINKS, Henriette et al. Transapical access closure : the TA PLUG device. In: Interactive cardiovascular and thoracic surgery, 2013, vol. 17, n° 5, p. 806–809; discussion 809–810. doi: 10.1093/icvts/ivt309
Main files (1)
Article (Published version)
accessLevelPublic
Identifiers
Journal ISSN1569-9285
10views
22downloads

Technical informations

Creation26/12/2025 15:21:38
First validation22/01/2026 09:19:24
Update22/01/2026 09:19:24
Status update22/01/2026 09:19:24
Last indexation22/01/2026 09:19:25
All rights reserved by Archive ouverte UNIGE and the University of GenevaunigeBlack