Scientific article
OA Policy
English

The self-expanding Symetis Acurate does not increase cerebral microembolic load when compared to the balloon-expandable Edwards Sapien prosthesis : a transcranial Doppler study in patients undergoing transapical aortic valve implantation

Published inPloS one, vol. 9, no. 10, e108191
Publication date2014
First online date2014-10-07
Abstract

Objectives: The aim of this study was to quantify potential differences in count, frequency and pattern of high-intensity transient signals (HITS) during transapical transcatheter aortic valve implantation (TA-TAVI), by comparing the Symetis Acurate TA (SA) with the balloon-expandable Edwards Sapien XT (ES) system.

Background: Recently, the Symetis Acurate TA revalving system has been introduced for TA-TAVI. The Symetis Acurate TA aortic bioprosthesis is self-expanding and is deployed by a specific two-step implantation technique. Whether this novel method increases the load of intraprocedural emboli, detected by transcranial Doppler ultrasound (TCD) as HITS, or not is not clear.

Methods: Twenty-two patients (n = 11 in each study arm, median logistic EuroScore 20%, median STS score 7%) displayed continuous TCD signals of good quality throughout the entire TA-TAVI procedure and were included in the final analysis. Data are presented as median with interquartile ranges.

Results: No significant differences were detected in total procedural or interval-related HITS load (SA: 303 [200; 594], ES: 499 [285; 941]; p = 0.16). With both devices, HITS peaked during prosthesis deployment (PD), whereas significantly fewer HITS occurred during instrumentation (SA: p = 0.002; ES: < 0.001) or post-implantation PI (SA: p = 0.007; ES: < 0.001). PD-associated HITS amounted to almost half of the total HITS load. One patient suffered new disabling stroke at 30 days. Thirty-day mortality amounted to 13.6% (3 of 22 patients).

Conclusions: Simplified transapical delivery using the self-expanding SA device does not increase HITS, despite of a two-step deployment technique with more interactions with the native aortic valve, when compared to the balloon-expandable ES valve. The similarity in HITS count, frequency and pattern with the two systems suggests a common mechanism for the release of cerebral microemboli.

Keywords
  • Aged
  • Aged, 80 and over
  • Female
  • Heart Valve Prosthesis Implantation / adverse effects
  • Heart Valve Prosthesis Implantation / methods
  • Humans
  • Intracranial Embolism / diagnosis
  • Intracranial Embolism / etiology
  • Male
  • Time Factors
  • Ultrasonography, Doppler, Transcranial / methods
Affiliation entities Not a UNIGE publication
Citation (ISO format)
ERDOES, Gabor et al. The self-expanding Symetis Acurate does not increase cerebral microembolic load when compared to the balloon-expandable Edwards Sapien prosthesis : a transcranial Doppler study in patients undergoing transapical aortic valve implantation. In: PloS one, 2014, vol. 9, n° 10, p. e108191. doi: 10.1371/journal.pone.0108191
Main files (1)
Article (Published version)
Identifiers
Journal ISSN1932-6203
10views
4downloads

Technical informations

Creation26/12/2025 15:21:16
First validation22/01/2026 08:46:22
Update22/01/2026 08:46:22
Status update22/01/2026 08:46:22
Last indexation22/01/2026 08:46:23
All rights reserved by Archive ouverte UNIGE and the University of GenevaunigeBlack