Scientific article
OA Policy
English

Novel emboli protection system during cardiac surgery : a multi-center, randomized, clinical trial

Published inThe annals of thoracic surgery, vol. 98, no. 5, p. 1627-1633; discussion 1633-1634
Publication date2014-11
First online date2014-09-23
Abstract

Background: Stroke is a major cause of morbidity and mortality during open-heart surgery. Up to 60% of intraoperative cerebral events are emboli induced. This randomized, controlled, multicenter trial is the first human study evaluating the safety and efficacy of a novel aortic cannula producing simultaneous forward flow and backward suction for extracting solid and gaseous emboli from the ascending aorta and aortic arch upon their intraoperative release.

Methods: Sixty-six patients (25 females; 68±10 years) undergoing elective aortic valve replacement surgery, with or without coronary artery bypass graft surgery, were randomized to the use of the CardioGard (CardioGard Medical, Or-Yehuda, Israel) Emboli Protection cannula ("treatment") or a standard ("control") aortic cannula. The primary endpoint was the volume of new brain lesions measured by diffusion-weighted magnetic resonance imaging (DW-MRI), performed preoperatively and postoperatively. Device safety was investigated by comparisons of complications rate, namely neurologic events, stroke, renal insufficiency and death.

Results: Of 66 patients (34 in the treatment group), 51 completed the presurgery and postsurgery MRI (27 in the treatment group). The volume of new brain lesion for the treatment group was (mean±standard error of the mean) 44.00±64.00 versus 126.56±28.74 mm3 in the control group (p=0.004). Of the treatment group, 41% demonstrated new postoperative lesions versus 66% in the control group (p=0.03). The complication rate was comparable in both groups.

Conclusions: The CardioGard cannula is safe and efficient in use during open-heart surgery. Efficacy was demonstrated by the removal of a substantial amount of emboli, a significant reduction in the volume of new brain lesions, and the percentage of patients experiencing new brain lesions.

Keywords
  • Aged
  • Aged, 80 and over
  • Aortic Valve Stenosis / surgery
  • Diffusion Magnetic Resonance Imaging
  • Embolic Protection Devices
  • Embolism / diagnosis
  • Embolism / epidemiology
  • Embolism / prevention & control
  • Equipment Design
  • Female
  • Follow-Up Studies
  • Heart Valve Prosthesis Implantation / adverse effects
  • Humans
  • Male
  • Middle Aged
  • Prospective Studies
  • Treatment Outcome
Affiliation entities Not a UNIGE publication
Citation (ISO format)
BOLOTIN, Gil et al. Novel emboli protection system during cardiac surgery : a multi-center, randomized, clinical trial. In: The annals of thoracic surgery, 2014, vol. 98, n° 5, p. 1627–1633; discussion 1633–1634. doi: 10.1016/j.athoracsur.2014.06.061
Main files (1)
Article (Published version)
accessLevelPublic
Identifiers
Journal ISSN0003-4975
10views
10downloads

Technical informations

Creation26/12/2025 15:21:20
First validation22/01/2026 08:42:03
Update24/03/2026 09:56:49
Status update24/03/2026 09:56:49
Last indexation24/03/2026 09:59:38
All rights reserved by Archive ouverte UNIGE and the University of GenevaunigeBlack