Abstract

Objectives: To validate a self-expanding transcatheter valve for off-pump transatrial mitral valve-in-ring (VIR) implantation via a left thoracotomy.

Methods: Mitral valve annuloplasty was performed via sternotomy during cardiopulmonary bypass on 9 pigs. After successful weaning from extracorporal circulation, the custom-made, self-expanding transcatheter VIR device was deployed under fluoroscopic guidance within the annuloplasty ring via a left thoracotomy. Hemodynamic data before and after the implantation were recorded. Mitral annulus diameter and valve area were measured by echocardiography. Transvalvular and left-ventricular outflow-tract pressure gradient were measured invasively.

Results: Eight successful implantations were performed. Implantation failed in 1 pig because of difficulty with technical delivery of the sheath. Mean transatrial procedure time was 12.6 ± 1.7 min. Hemodynamic status during transatrial implantation was stable, and differences were not statistically significant. Mean mitral annulus diameter and mean mitral orifice area were 2.32 ± 0.2 and 3.84 ± 0.55 cm2, respectively. Mild regurgitation was detected in 7 animals and moderate regurgitation in 1. Mean gradients were 6.1 ± 5.0 mm Hg across the device. Postmortem examination confirmed adequate positioning of devices within the annuloplasty ring.

Conclusions: This custom-made transcatheter device allows for safe and reproducible off-pump transatrial mitral VIR implantations. Transatrial access is a promising route to facilitate VIR implantations. Our custom-made stent-valve may be suitable for VIR procedures.

Keywords
  • Animals
  • Cardiac Catheterization / methods
  • Cardiopulmonary Bypass
  • Echocardiography
  • Heart Valve Prosthesis Implantation / instrumentation
  • Hemodynamics
  • Mitral Valve / diagnostic imaging
  • Mitral Valve / surgery
  • Mitral Valve Annuloplasty / instrumentation
  • Operative Time
  • Prosthesis Design
  • Prosthesis Failure
  • Stents
  • Swine
  • Thoracotomy
Affiliation entities Not a UNIGE publication
Citation (ISO format)
MA, Liang et al. New Self-Expanding Transcatheter Valve for Off-Pump Transatrial Mitral Valve-In-Ring Implantation. In: Cardiology, 2015, vol. 132, n° 4, p. 221–227. doi: 10.1159/000433595
Main files (1)
Article (Published version)
accessLevelRestricted
Identifiers
Journal ISSN0008-6312
10views
0downloads

Technical informations

Creation26/12/2025 15:21:07
First validation21/01/2026 16:44:09
Update21/01/2026 16:44:09
Status update21/01/2026 16:44:09
Last indexation21/01/2026 16:44:10
All rights reserved by Archive ouverte UNIGE and the University of GenevaunigeBlack