Rupture of a polypropylene suture after aortic operation : a scanning electronic microscopical assessment of potential mechanisms

Published inThe annals of thoracic surgery, vol. 75, no. 4, p. 1318-1321
Publication date2003-04
Abstract

In surgical procedures, and specifically in cardiovascular operations, the polypropylene monofilament is a widely used suture material. We report an unexplained case of early rupture of a 4-0 polypropylene suture used for the distal anastomosis after replacement of the ascending aorta with a Dacron (Vaskutek, Sulzer Medica, Austin, TX) prosthesis. By trying to identify if intraoperative mechanical damage could have led to the rupture, we obtained impressive results of how different manipulations of suture material using surgical instruments can damage those delicate filaments. Pictures were obtained by light microscopy and electron microscopy.

Keywords
  • Aorta / surgery
  • Aortic Aneurysm / surgery
  • Female
  • Humans
  • Middle Aged
  • Polypropylenes
  • Reoperation
  • Stress, Mechanical
  • Sutures
Affiliation entities Not a UNIGE publication
Citation (ISO format)
HUBER, Christoph et al. Rupture of a polypropylene suture after aortic operation : a scanning electronic microscopical assessment of potential mechanisms. In: The annals of thoracic surgery, 2003, vol. 75, n° 4, p. 1318–1321. doi: 10.1016/s0003-4975(02)04679-9
Main files (1)
Article (Published version)
Identifiers
Journal ISSN0003-4975
9views
61downloads

Technical informations

Creation26/12/2025 15:23:12
First validation21/01/2026 16:01:17
Update21/01/2026 16:01:17
Status update21/01/2026 16:01:17
Last indexation21/01/2026 16:01:18
All rights reserved by Archive ouverte UNIGE and the University of GenevaunigeBlack