From mouth-to-mouth to bag-valve-mask ventilation : evolution and characteristics of actual devices--a review of the literature

Published inBioMed research international, vol. 2014, 762053
Publication date2014
First online date2014-05-27
Abstract

Manual ventilation is a vital procedure, which remains difficult to achieve for patients who require ventilatory support. It has to be performed by experienced healthcare providers that are regularly trained for the use of bag-valve-mask (BVM) in emergency situations. We will give in this paper, a historical view on manual ventilation's evolution throughout the last decades and describe the technical characteristics, advantages, and hazards of the main devices currently found in the market. Artificial ventilation has developed progressively and research is still going on to improve the actual devices used. Throughout the past years, a brand-new generation of ventilators was developed, but little was done for manual ventilation. Many adverse outcomes due to faulty valve or misassembly were reported in the literature, as well as some difficulties to ensure efficient insufflation according to usual respiratory parameters. These serious incidents underline the importance of BVM system routine check and especially the unidirectional valve reassembly after sterilization, by only experienced and trained personnel. Single use built-in devices may prevent disassembly problems and are safer than the reusable ones. Through new devices and technical improvements, the safety of BVM might be increased.

Keywords
  • Cardiopulmonary Resuscitation / adverse effects
  • Cardiopulmonary Resuscitation / history
  • Cardiopulmonary Resuscitation / instrumentation
  • History, 15th Century
  • History, 16th Century
  • History, 18th Century
  • History, 19th Century
  • History, 20th Century
  • Humans
  • Respiration, Artificial / adverse effects
  • Respiration, Artificial / history
  • Respiration, Artificial / instrumentation
Affiliation entities Not a UNIGE publication
Citation (ISO format)
KHOURY, Abdo et al. From mouth-to-mouth to bag-valve-mask ventilation : evolution and characteristics of actual devices--a review of the literature. In: BioMed research international, 2014, vol. 2014, p. 762053. doi: 10.1155/2014/762053
Main files (1)
Article (Published version)
Identifiers
Journal ISSN2314-6133
10views
31downloads

Technical informations

Creation23/12/2025 14:15:54
First validation21/01/2026 15:44:58
Update21/01/2026 15:44:58
Status update21/01/2026 15:44:58
Last indexation21/01/2026 15:45:08
All rights reserved by Archive ouverte UNIGE and the University of GenevaunigeBlack