Scientific article
OA Policy
English

Evaluation of end-to-end 3D absorbed dose distribution in 90Y-SIRT and SBRT combination therapy using MAGIC-f polymer gel dosimeter

Published inEuropean journal of nuclear medicine and molecular imaging, vol. 53, no. 2, p. 1313-1330
First online date2026
Abstract

Background: Selective Internal Radiation Therapy (SIRT) with 90 Y-microspheres, and Stereotactic Body Radiation Therapy (SBRT), are well-established treatment strategies for liver malignancies. Combining these two modalities has proven to be an effective and safe approach for addressing undertreated tumor regions from initial SIRT through a boost dose from SBRT. The complex dosimetry process, which includes image-based dosimetry of 90 Y-SIRT and SBRT dose painting, requires precise dosimetry verification. In this study, gel dosimetry is proposed as a valuable novel tool to capture the 3D absorbed dose contributions from both treatments with high spatial resolution.

Methods: The MAGIC-f polymer gel dosimeter was used in three experiments: external beam radiation therapy (EBRT), 90 Y-SIRT, and combination therapy. For EBRT, eight calibration vials filled with gel were irradiated with absorbed doses ranging from 0 to 21 Gy. In the 90 Y-SIRT experiments, eleven vials were filled with gel containing varying concentrations of 90 Y-citrate, delivering doses from 0 to 44 Gy after 68 h. For combination therapy, eight vials received 0-7 Gy from 90 Y-citrate after 68 h, followed by an additional 0-7 Gy from SBRT (final doses 0-14 Gy). A phantom, designed to simulate a tumor within a normal liver, was created with the sphere receiving 10 Gy after 68 h from SIRT and another 10 Gy from SBRT using 6MV photons (in total 20 Gy). PET/CT imaging was conducted before SBRT, and the Medical Internal Radiation Dose (MIRD) schema was used for dose calculations. For each experiment, MR T2-weighted imaging was performed using a 3 Tesla scanner, and R2 maps (1/s) were analyzed to establish a dose-response relationship between absorbed dose (Gy) and R2. The gel's sensitivity to each irradiation was measured, and phantom's dose maps were evaluated using mean absorbed dose, dose volume histograms (DVHs), line profiles, and isodose maps.

Results: The dose response of the gel was linear within the irradiated ranges for EBRT and combination therapy. The linear range for 90 Y-SIRT was between 0 and 16.75 Gy. The sensitivity of the gel was 0.380, 0.758, and 0.713 s⁻¹/Gy for EBRT, 90 Y-SIRT, and combination experiments, respectively. In the phantom irradiated with combination therapy, a saturated area was observed in the central core of the sphere, surrounded by an underestimation area with a diameter of 4.5 mm. The mean absorbed dose values within the sphere were 9.83 Gy, 9.71 Gy, and 18.58 Gy from SBRT, SIRT, and combination therapy, respectively. For the cylinder, these values were 1.29 Gy, 0.61 Gy, and 2.68 Gy, respectively. The DVHs, line profiles, and isodose lines for the combination therapy demonstrated the cumulative effects of the absorbed dose from both treatments.

Conclusion: This is the first study demonstrating the feasibility of using MAGIC-f gel dosimetry to directly measure 3D dose distributions from combined 90 Y-SIRT and SBRT. Compared to PET/CT-based dosimetry, the gel method offers superior spatial resolution and enables objective physical verification of complex dose distributions. This technique has strong potential for quality assurance in theranostic protocols, particularly for beta emitters and, potentially, alpha-emitting radiotracers in future applications.

Keywords
  • 90Y-SIRT
  • 3D dosimetry
  • Combination therapy
  • Gel dosimetry
  • MAGIC-f
  • SBRT
Citation (ISO format)
MANSOURI, Zahra et al. Evaluation of end-to-end 3D absorbed dose distribution in 90Y-SIRT and SBRT combination therapy using MAGIC-f polymer gel dosimeter. In: European journal of nuclear medicine and molecular imaging, 2026, vol. 53, n° 2, p. 1313–1330. doi: 10.1007/s00259-025-07461-2
Main files (1)
Article (Published version)
Secondary files (1)
Supplemental data
accessLevelPublic
Identifiers
Journal ISSN1619-7070
27views
180downloads

Technical informations

Creation08/08/2025 20:17:46
First validation14/08/2025 06:55:23
Update10/02/2026 10:32:40
Status update10/02/2026 10:32:40
Last indexation10/02/2026 10:32:41
All rights reserved by Archive ouverte UNIGE and the University of GenevaunigeBlack