Scientific article
Case report
English

Variations in autoregulation-based optimal cerebral perfusion pressure determination using two integrated neuromonitoring platforms in a trauma patient

Published inNeurocritical care, vol. 41, no. 2, p. 386-392
Publication date2024-10
First online date2024-02-29
Abstract

Background: Neuromonitoring devices are often used in traumatic brain injury. The objective of this report is to raise awareness concerning variations in optimal cerebral perfusion pressure (CPPopt) determination using exploratory information provided by two neuromonitoring monitors that are part of research programs (Moberg CNS Monitor and RAUMED NeuroSmart LogO).

Methods: We connected both monitors simultaneously to a parenchymal intracranial pressure catheter and recorded the pressure reactivity index (PRx) and the derived CPPopt estimates for a patient with a severe traumatic brain injury. These estimates were available at the bedside and were updated at each minute.

Results: Using the Bland and Altman method, we found a mean variation of - 3.8 (95% confidence internal from - 8.5 to 0.9) mm Hg between the CPPopt estimates provided by the two monitors (limits of agreement from - 26.6 to 19.1 mm Hg). The PRx and CPPopt trends provided by the two monitors were similar over time, but CPPopt trends differed when PRx values were around zero. Also, almost half of the CPPopt estimates differed by more than 10 mm Hg.

Conclusions: These wide variations recorded in the same patient are worrisome and reiterate the importance of understanding and standardizing the methodology and algorithms behind commercial neuromonitoring devices prior to incorporating them in clinical use.

Keywords
  • Cerebral autoregulation
  • Cerebral perfusion pressure
  • Neuromonitoring
  • Optimal cerebral perfusion pressure
  • Pressure reactivity index
  • Traumatic brain injury
Citation (ISO format)
PLOURDE, Guillaume et al. Variations in autoregulation-based optimal cerebral perfusion pressure determination using two integrated neuromonitoring platforms in a trauma patient. In: Neurocritical care, 2024, vol. 41, n° 2, p. 386–392. doi: 10.1007/s12028-024-01949-9
Main files (1)
Article (Published version)
accessLevelRestricted
Secondary files (2)
Supplemental data
accessLevelRestricted
Supplemental data
accessLevelRestricted
Identifiers
Additional URL for this publicationhttps://link.springer.com/10.1007/s12028-024-01949-9
Journal ISSN1541-6933
23views
0downloads

Technical informations

Creation23/11/2024 15:37:01
First validation16/07/2025 09:32:05
Update13/10/2025 11:31:38
Status update16/07/2025 09:32:05
Last indexation12/11/2025 16:54:24
All rights reserved by Archive ouverte UNIGE and the University of GenevaunigeBlack