Intradevice Repeatability and Interdevice Comparison of Two Specular Microscopy Devices in a Real-Life Setting: Tomey EM-4000 and Nidek CEM-530

Published inMedicina, vol. 60, no. 7, 1110
Publication date2024-07-09
First online date2024-07-09
Abstract

Background and Objectives: The purpose of this study was to compare two commercially available specular microscopes (Tomey EM-4000 and Nidek CEM-530) in a real-life clinical setting in terms of intra- and interdevice variability. The study was conducted on all patients seen in a clinical practice specializing in anterior segment pathologies, regardless of the purpose of their visit.

Materials and Methods: In total, 112 eyes of 56 patients (age 23–85 years old) were included in the study. Each eye was measured three times with each device (for a total of six measurements), and results for central corneal thickness (CCT) and corneal endothelial cell density (ECD) were recorded. The results were then evaluated with the D’Agostino–Pearson normality test and compared with a Wilcoxon signed-rank test, t-test, ANOVA or Mann–Whitney test for intra- and interdevice variability.

Results: Both specular microscopes produced very reliable reproducible intradevice results: The Tomey EM-4000 measured an ECD of 2390 ± 49.57 cells/mm2 (mean ± standard error of mean); the range was 799–3010 cells/mm2. The determined CCT was 546 ± 5.104 µm (mean ± standard error of mean [SEM]); the range was 425–615 µm. The measurements with the Nidek CEM-530 revealed an ECD of 2417 ± 0.09 cells/mm2 (mean ± SEM); the range was 505–3461 cells/mm2 (mean ± SEM). The mean CCT detected was 546.3 ± 4.937 µm (mean ± SEM); the range was 431–621 µm. The interdevice differences were statistically significant for both parameters, ECD (p = 0.0175) and CCT (p = 0.0125) (p < 0.05).

Conclusions: The Nidek CEM-530 and the Tomey EM-4000 both produced reliable and reproducible results in terms of ECD and CCT. The absolute measurements were statistically significantly different for CCT and ECD for both devices; the Nidek produces slightly higher values.

Keywords
  • Fuchs’ dystrophy
  • Anterior segment
  • Corneal endothelium
  • Specular microscopy
  • Adult
  • Aged
  • Aged, 80 and over
  • Cell Count / instrumentation
  • Cell Count / methods
  • Female
  • Humans
  • Male
  • Microscopy / instrumentation
  • Microscopy / methods
  • Middle Aged
  • Reproducibility of Results
  • Young Adult
Citation (ISO format)
KECIK, Mateusz Mariusz et al. Intradevice Repeatability and Interdevice Comparison of Two Specular Microscopy Devices in a Real-Life Setting: Tomey EM-4000 and Nidek CEM-530. In: Medicina, 2024, vol. 60, n° 7, p. 1110. doi: 10.3390/medicina60071110
Main files (1)
Article (Published version)
Secondary files (5)
Identifiers
Additional URL for this publicationhttps://www.mdpi.com/1648-9144/60/7/1110
Journal ISSN1010-660X
11views
2downloads

Technical informations

Creation20/10/2024 08:21:01
First validation19/03/2025 08:06:00
Update time19/03/2025 08:06:00
Status update19/03/2025 08:06:00
Last indexation19/03/2025 08:06:01
All rights reserved by Archive ouverte UNIGE and the University of GenevaunigeBlack