Accuracy of surgical guides manufactured with four different 3D printers. A comparative in vitro study

Published inJournal of dentistry, vol. 148, 105226
Publication date2024-09
First online date2024-07-04
Abstract

Objectives: The aim of this study was to assess the accuracy of surgical guides manufactured with four different 3D printers..

Methods: Forty-eight surgical guides (BlueSky Plan, BlueSky Bio) were produced using four different 3D printers, with strict adherence to each manufacturer's instructions. The printers used were three digital light processing (DLP) printers (SolFlex170, VC; Nextdent5100, ND, and D30+Rapidshape, RS) and one stereolithographic (SLA) printer (Formlabs3B+, FL). The study evaluated the trueness and precision of the overall surface, the region of interest (RoI) (occlusal and guide zone), the repeatability in several batches, and the guide hole's diameter and xyz axes. The printed guides were digitized and compared with the CAD design control specimen (Control X, Geomagic). Descriptive statistics and Kruskal-Wallis tests with post-hoc Mann-Whitney tests were performed (α=0.05).

Results: Differences in trueness and precision were found between groups in the overall zone and RoI (p = 0.00). The ND group demonstrated the highest repeatability. Only the RS group exhibited a comparable guide hole diameter to the master specimen (5.27±2.12 mm; p = 0.104). No statistical differences were observed between groups in the x and z axes. However, in the y-axis, the VC group displayed statistically significant differences (p = 0.01).

Conclusions: The results showed that the DLP groups had better overall accuracy, while the SLA group had the best results in the RoI. The manufacturer's workflows demonstrated a high reproducibility between batches in the RoI. The RS group had values most similar values to the guide hole diameter of the master specimen, with minimal deviations in guide hole orientation.

Clinical significance: Implant position can be affected by the accuracy of the 3D printed surgical guide. Therefore, it is critical to analyze the final dimensions and the direction of the guide hole using available printing technologies.

Keywords
  • 3D printing
  • Accuracy
  • Guided implant surgery
  • Implantology
  • Stereolithography
  • Surgical guide
  • Computer-Aided Design
  • Equipment Design
  • Humans
  • Imaging, Three-Dimensional
  • Models, Dental
  • Printing, Three-Dimensional
  • Reproducibility of Results
  • Surgery, Computer-Assisted / instrumentation
  • Surgery, Computer-Assisted / methods
  • Comparative Study
Citation (ISO format)
MORÓN-CONEJO, Belén et al. Accuracy of surgical guides manufactured with four different 3D printers. A comparative in vitro study. In: Journal of dentistry, 2024, vol. 148, p. 105226. doi: 10.1016/j.jdent.2024.105226
Main files (1)
Article (Published version)
Identifiers
Journal ISSN0300-5712
48views
488downloads

Technical informations

Creation22/01/2025 16:56:55
First validation29/01/2025 16:19:27
Update29/01/2025 16:19:27
Status update29/01/2025 16:19:27
Last indexation29/01/2025 16:19:28
All rights reserved by Archive ouverte UNIGE and the University of GenevaunigeBlack