Scientific article
English

The Accuracy of Full-Arch Intraoral Optical Impressions (IOS): Clinical Pilot Study of the Influence of the Scan Strategy, Operator, and Intraoral Scanner

Published inThe International journal of prosthodontics, vol. 36, no. 6, p. 689-696
Publication date2023-12-18
First online date2023-12-18
Abstract

Purpose: To investigate the influence of a novel scanning strategy-using two new intraoral scanner devices with different operators-on the full-arch scanning accuracy for a dentate maxilla.

Materials and methods: Two scanning strategies, a test and a control strategy, were used to produce full-arch impressions of the dentate maxilla of a study patient. Two intraoral scanning (IOS) devices were used. Five expert operators performed a total of 40 scans. The scan time was recorded for each. A reference model was obtained from the patient's maxillary arch with an analog impression. The model was later scanned with a high-precision laboratory scanner to create a digital reference model (DRM). The scanning accuracy was analyzed with 3D-analysis software using a root mean square (RMS) calculation method, and qualitative analysis was executed using machine learning software.

Results: The mean RMS result for the test strategy was 82.8 ± 16 μm compared to 81.5 ± 16 μm for the control strategy. The mean RMS results were 84.7 ± 15 μm for Primescan (PS) and 79.6 ± 17 μm for 3Shape (3S). As such, the scanning strategies and IOS devices did not influence the scanning accuracy. Yet, a significant difference was found when the two strategies' scanning times were compared (P = .001), as well as the IOS devices (P = .001). The operator was found to have no influence on the scanning strategy.

Conclusions: The accuracy of digital impressions is not influenced by different strategies, devices, or operators, in contrast with the scanning time, which is influenced by both strategies and devices.

Keywords
  • Computer-Aided Design
  • Dental Arch
  • Dental Impression Technique
  • Humans
  • Imaging, Three-Dimensional
  • Models, Dental
  • Pilot Projects
Citation (ISO format)
CORDARO, Matteo et al. The Accuracy of Full-Arch Intraoral Optical Impressions (IOS): Clinical Pilot Study of the Influence of the Scan Strategy, Operator, and Intraoral Scanner. In: The International journal of prosthodontics, 2023, vol. 36, n° 6, p. 689–696. doi: 10.11607/ijp.8113
Main files (1)
Article (Published version)
accessLevelRestricted
Identifiers
Journal ISSN0893-2174
42views