Continuous cardiac output monitoring after cardiopulmonary bypass : a comparison with bolus thermodilution measurement

Published inIntensive care medicine, vol. 32, no. 6, p. 919-922
Publication date2006-06
First online date2006-04-07
Abstract

Objective: The interchangeability of continuous measurement of cardiac output (CO) with the traditional bolus method in patients after cardiopulmonary bypass (CPB) is uncertain.

Design: Prospective observational clinical study.

Setting: A 20-bed surgical ICU at a university hospital.

Patients: Fourteen deeply sedated, ventilated, post-cardiac surgery patients, all equipped with a pulmonary artery catheter.

Interventions: Six hours after the end of the CPB, 56 simultaneous bolus and continuous measurements were compared by a linear regression analysis and Bland-Altman analysis. Bolus CO was estimated by averaging triplicate injections of 10 ml room-temperature NaCl 0.9%, delivered randomly during the respiratory cycle. A stringent maximum difference of 0.55 l min(-1) (about 10% of the mean bolus measured) was considered as a clinically acceptable agreement between the two types of measurements. To be interchangeable the limits of agreement (+/-2 SD of the mean difference between the two methods) should not exceed the chosen acceptable difference.

Measurements and results: Continuous was correlated with bolus CO, with a correlation coefficient of r(2)=0.68. (p<0.01). The Bland-Altman analysis demonstrated an objective mean bias of 0.33+/-0.6 l min(-1) (confidence interval of -0.87-1.58) with 34% of measured values falling outside of the clinically acceptable limits.

Conclusion: Our results suggest that, in the first 6 h after CPB, continuous and bolus CO determinations are not interchangeable; one third of the values obtained by continuous CO fell outside the strict limits of clinically useful precision.

Keywords
  • Aged
  • Cardiac Output / physiology
  • Cardiopulmonary Bypass
  • Catheterization, Swan-Ganz
  • Female
  • Humans
  • Hypothermia
  • Linear Models
  • Male
  • Middle Aged
  • Monitoring, Physiologic
  • Prospective Studies
  • Thermodilution
  • Thoracic Surgery
Citation (ISO format)
BENDJELID, Karim et al. Continuous cardiac output monitoring after cardiopulmonary bypass : a comparison with bolus thermodilution measurement. In: Intensive care medicine, 2006, vol. 32, n° 6, p. 919–922. doi: 10.1007/s00134-006-0161-2
Main files (1)
Article (Published version)
accessLevelRestricted
Identifiers
Journal ISSN0342-4642
33views
0downloads

Technical informations

Creation20/11/2024 15:35:53
First validation14/01/2025 08:39:36
Update14/01/2025 08:39:36
Status update14/01/2025 08:39:36
Last indexation14/01/2025 08:39:37
All rights reserved by Archive ouverte UNIGE and the University of GenevaunigeBlack