Scientific article
English

Hybrid alumina total hip arthroplasty using a press-fit metal-backed socket in patients younger than 55 years. A six- to 11-year evaluation

Published inJournal of bone and joint surgery. British volume, vol. 86, no. 2, p. 190-194
Publication date2004-03
Abstract

Between 1990 and 1992, we implanted 71 hybrid alumina-on-alumina hip arthroplasties in 62 consecutive patients under the age of 55 years, with a mean age of 46 years at surgery. There were 56 primary and 15 secondary procedures. The prostheses involved a cemented titanium alloy stem, a 32 mm alumina head, and a press-fit metal-backed socket with an alumina insert. Three patients (four hips) died from unrelated causes. Four hips had revision surgery for either deep infection, unexplained persistent pain, fracture of the alumina head, or aseptic loosening of the socket. The nine-year survival rate was 93.7% with revision for any cause as the end-point and 98.4% with revision for aseptic loosening as the end-point. The outcome in the surviving patients (50 patients, 57 hips) with a minimum five-year follow-up (mean eight years) was excellent in 47 hips (82.5%), very good in eight (14%), good in one and fair in one. A thin, partial, lucent line, mainly in zone III was present in 38% of the sockets and one socket had a complete lucency less than 1 mm thick. One stem had isolated femoral osteolysis. There was no detectable component migration nor acetabular osteolysis. This hybrid arthroplasty gave satisfactory medium-term results in active patients. The press-fit metal-backed socket appeared to have reliable fixation in alumina-on-alumina hip arthroplasty. The excellent results using cemented fixation of the stem may be related to the low production of wear debris.

Keywords
  • Adult
  • Aluminum Oxide
  • Arthroplasty, Replacement, Hip / methods
  • Arthroplasty, Replacement, Hip / mortality
  • Cementation
  • Female
  • Femur Head Necrosis / surgery
  • Hip Prosthesis / standards
  • Humans
  • Male
  • Middle Aged
  • Osteoarthritis, Hip / surgery
  • Pain, Postoperative / etiology
  • Pain, Postoperative / surgery
  • Prosthesis Design / standards
  • Prosthesis Failure
  • Prosthesis-Related Infections / etiology
  • Prosthesis-Related Infections / mortality
  • Reoperation
  • Treatment Outcome
Affiliation entities Not a UNIGE publication
Citation (ISO format)
BIZOT, Pascal et al. Hybrid alumina total hip arthroplasty using a press-fit metal-backed socket in patients younger than 55 years. A six- to 11-year evaluation. In: Journal of bone and joint surgery. British volume, 2004, vol. 86, n° 2, p. 190–194. doi: 10.1302/0301-620x.86b2.14026
Main files (1)
Article (Published version)
accessLevelRestricted
Identifiers
Journal ISSN0301-620X
57views
0downloads

Technical informations

Creation17/11/2024 19:14:31
First validation19/11/2024 10:03:14
Update30/04/2026 10:45:32
Status update30/04/2026 10:45:32
Last indexation30/04/2026 10:47:33