Safety and performance of the third-generation drug-eluting resorbable coronary magnesium scaffold system in the treatment of subjects with de novo coronary artery lesions : 6-month results of the prospective, multicenter BIOMAG-I first-in-human study

Published inEClinicalMedicine, vol. 59, 101940
Publication date2023-05
First online date2023-04-17
Abstract

Background: A third-generation coronary drug-eluting resorbable magnesium scaffold (DREAMS 3G) was developed to enhance the performance of previous scaffold generations and achieve angiographic outcomes comparable to those of contemporary drug-eluting stents.

Methods: This prospective, multicenter, non-randomized, first-in-human study was conducted at 14 centers in Europe. Eligible patients had stable or unstable angina, documented silent ischemia, or non-ST-elevation myocardial infarction, and a maximum of two single de novo lesions in two separate coronary arteries with a reference vessel diameter between 2.5 mm and 4.2 mm. Clinical follow-up was scheduled at one, six and 12 months and annually thereafter until five years. Invasive imaging assessments were scheduled six and 12 months postoperatively. The primary endpoint was angiographic in-scaffold late lumen loss at six months. This trial was registered at ClinicalTrials.gov (NCT04157153).

Findings: Between April 2020 and February 2022, 116 patients with 117 coronary artery lesions were enrolled. At six months, in-scaffold late lumen loss was 0.21 mm (SD 0.31). Intravascular ultrasound assessment showed preservation of the scaffold area (mean 7.59 mm2[SD 2.21] post-procedure vs 6.96 mm2[SD 2.48]) at six months) with a low mean neointimal area (0.02 mm2[SD 0.10]). Optical coherence tomography revealed that struts were embedded in the vessel wall and were already hardly discernible at six months. Target lesion failure occurred in one (0.9%) patient; a clinically driven target lesion revascularization was performed on post-procedure day 166. No definite or probable scaffold thrombosis or myocardial infarction was observed.

Interpretation: These findings show that the implantation of DREAMS 3G in de novo coronary lesions is associated with favorable safety and performance outcomes, comparable to contemporary drug-eluting stents.

eng
Keywords
  • Bioresobabe scaffolds
  • Coronary artery disease
  • Drug-eluting stents
Citation (ISO format)
HAUDE, Michael et al. Safety and performance of the third-generation drug-eluting resorbable coronary magnesium scaffold system in the treatment of subjects with de novo coronary artery lesions : 6-month results of the prospective, multicenter BIOMAG-I first-in-human study. In: EClinicalMedicine, 2023, vol. 59, p. 101940. doi: 10.1016/j.eclinm.2023.101940
Main files (1)
Article (Published version)
Secondary files (2)
Identifiers
ISSN of the journal2589-5370
12views
1downloads

Technical informations

Creation07/04/2024 8:15:38 AM
First validation07/23/2024 8:49:49 AM
Update time07/23/2024 8:49:49 AM
Status update07/23/2024 8:49:49 AM
Last indexation07/23/2024 8:50:10 AM
All rights reserved by Archive ouverte UNIGE and the University of GenevaunigeBlack