en
Scientific article
Case report
English

Intraocular pressure variations after intravitreal injections measured with an implanted suprachoroidal telemetry sensor

Published inJournal of glaucoma, vol. 30, no. 8, p. e360-e363
Publication date2021-08-12
First online date2021-05-12
Abstract

Purpose:

Intravitreal injections (IVIs) may create transient intraocular pressure (IOP) elevation. This report describes continuous IOP fluctuations following multiple IVI measured with a permanent implantable sensor.

Patients and Methods:

We report the case of a 49-year-old white glaucomatous male with refractory macular edema secondary to central retinal vein occlusion in his left eye who underwent deep sclerectomy combined with the implantation of a suprachoroidal tonometry sensor. Serial IOP measurements were performed immediately before and after each IVI over a 1-year period.

Results:

During the first 7 months following deep sclerectomy, IOP remained below 10 mm Hg. During this period, mean IOP before each injection was 2.1±2.6 mm Hg, and each IVI caused a reduction of 1.2±0.8 mm Hg on average, with a maximum reduction of 2.7 mm Hg, before IOP normalized within 50 minutes to 24 hours. From 7 months postoperatively, mean IOP increased to the low teens. During this period, mean IOP before each injection was 9.9±1.8 mm Hg, and each IVI caused an increase of 15.8±11.7 mm Hg on average, with a maximum increase of 44.8 mm Hg, before IOP normalized within 20 minutes to 4 hours.

Conclusions:

During the initial postoperative phase, IVI may cause acute reduction in IOP, either through subconjunctival leaks or increased filtration secondary to increased fluid pressure. Several months after surgery, this effect subsides and IOP spikes sharply immediately after each IVI, suggesting the resolution of the initial mechanism, most likely through scarring and fibrosis.

eng
Keywords
  • Glaucoma
  • Continuous
  • 24 hours
  • IOP
  • IVT
  • Aflibercept
  • VEGF
  • Intravitreal injections
  • Eyemate
Affiliation Not a UNIGE publication
Citation (ISO format)
MEDURI, Enrico, GILLMANN, Kevin, MANSOURI, Kaweh. Intraocular pressure variations after intravitreal injections measured with an implanted suprachoroidal telemetry sensor. In: Journal of glaucoma, 2021, vol. 30, n° 8, p. e360–e363. doi: 10.1097/IJG.0000000000001879
Main files (1)
Article (Published version)
accessLevelRestricted
Identifiers
ISSN of the journal1057-0829
21views
0downloads

Technical informations

Creation03/19/2024 11:10:06 AM
First validation05/14/2024 7:08:55 AM
Update time05/14/2024 7:08:55 AM
Status update05/14/2024 7:08:55 AM
Last indexation05/14/2024 7:09:16 AM
All rights reserved by Archive ouverte UNIGE and the University of GenevaunigeBlack