Scientific article
English

Unacceptable failure of the PI2® implant

Published inJournal of hand surgery. European volume, vol. 41, no. 9, p. 917-922
Publication date2016-11
First online date2016-09-28
Abstract

Unlabelled: The PI2® spacer is designed for treatment of trapeziometacarpal (TM) osteoarthritis. However, the shape of this implant has raised concerns about its stability. We retrospectively investigated 45 implants in 41 patients treated for trapeziometacarpal osteoarthritis in our hospital between 2004 and 2009 who underwent trapeziectomy and insertion of a PI2® spacer. Outcome parameters included revision rates and clinical outcomes correlated with implant position and scaphometacarpal distance, assessed using standard radiographs. A total of 12 implants (27%) were removed at a median time of 10 months (interquartile range (IQR), 7-22 months). These included five dislocations and one early infection. Additionally, a further six patients underwent revision due to persistent pain. Three of these had scapho-trapezoid osteoarthritis, two had developed subluxation of the implant, and one did not show any radiographic abnormalities. A review of patient records revealed that 33 implants remained in place at a median time of 29 months (IQR, 20-57). However, of those, only 21 implants (64%) in 17 patients were available for clinical evaluation at a median follow-up of 29 months (IQR, 19-62 months). No significant differences in clinical outcomes including functional results were observed between in-place ( n = 8) and subluxated ( n = 13) implants. Due to the high revision rate (12/45), consistent with other reports in the literature, we have abandoned the use of the PI2® spacer. We recommend the establishment of a registry for evaluation of future implants.

Type of study/level of evidence: Case-series study/level IV.

Keywords
  • PI2®
  • Osteoarthritis
  • Pyrocarbon
  • Trapeziometacarpal joint
  • Aged
  • Aged, 80 and over
  • Carpometacarpal Joints
  • Female
  • Hand Strength
  • Humans
  • Joint Prosthesis / adverse effects
  • Male
  • Middle Aged
  • Osteoarthritis / surgery
  • Patient Satisfaction
  • Prosthesis Design
  • Prosthesis Failure
  • Retrospective Studies
  • Trapezium Bone / surgery
  • Treatment Outcome
Citation (ISO format)
VAN AAKEN, Jan et al. Unacceptable failure of the PI2® implant. In: Journal of hand surgery. European volume, 2016, vol. 41, n° 9, p. 917–922. doi: 10.1177/1753193416651573
Main files (1)
Article (Published version)
accessLevelRestricted
Identifiers
Journal ISSN1753-1934
53views
0downloads

Technical informations

Creation03/08/2024 1:49:25 PM
First validation05/08/2024 3:01:29 PM
Update time05/08/2024 3:01:29 PM
Status update05/08/2024 3:01:29 PM
Last indexation11/01/2024 9:30:06 AM
All rights reserved by Archive ouverte UNIGE and the University of GenevaunigeBlack