Impact of simulated reduced injected dose on the assessment of amyloid PET scans

Published inEuropean journal of nuclear medicine and molecular imaging, vol. 51, no. 3, p. 734-748
Publication date2024-02
First online date2023-10-28
Abstract

Purpose: To investigate the impact of reduced injected doses on the quantitative and qualitative assessment of the amyloid PET tracers [18F]flutemetamol and [18F]florbetaben.

Methods: Cognitively impaired and unimpaired individuals (N = 250, 36% Aβ-positive) were included and injected with [18F]flutemetamol (N = 175) or [18F]florbetaben (N = 75). PET scans were acquired in list-mode (90-110 min post-injection) and reduced-dose images were simulated to generate images of 75, 50, 25, 12.5 and 5% of the original injected dose. Images were reconstructed using vendor-provided reconstruction tools and visually assessed for Aβ-pathology. SUVRs were calculated for a global cortical and three smaller regions using a cerebellar cortex reference tissue, and Centiloid was computed. Absolute and percentage differences in SUVR and CL were calculated between dose levels, and the ability to discriminate between Aβ- and Aβ + scans was evaluated using ROC analyses. Finally, intra-reader agreement between the reduced dose and 100% images was evaluated.

Results: At 5% injected dose, change in SUVR was 3.72% and 3.12%, with absolute change in Centiloid 3.35CL and 4.62CL, for [18F]flutemetamol and [18F]florbetaben, respectively. At 12.5% injected dose, percentage change in SUVR and absolute change in Centiloid were < 1.5%. AUCs for discriminating Aβ- from Aβ + scans were high (AUC ≥ 0.94) across dose levels, and visual assessment showed intra-reader agreement of > 80% for both tracers.

Conclusion: This proof-of-concept study showed that for both [18F]flutemetamol and [18F]florbetaben, adequate quantitative and qualitative assessments can be obtained at 12.5% of the original injected dose. However, decisions to reduce the injected dose should be made considering the specific clinical or research circumstances.

Keywords
  • Alzheimer’s disease
  • Amyloid
  • Dose reduction
  • Neuroimaging
  • PET
  • Alzheimer Disease / diagnostic imaging
  • Amyloid / metabolism
  • Amyloid beta-Peptides / metabolism
  • Aniline Compounds
  • Benzothiazoles
  • Brain / metabolism
  • Humans
  • Positron-Emission Tomography / methods
  • Stilbenes
  • Flutemetamol
  • 4-(N-methylamino)-4'-(2-(2-(2-fluoroethoxy)ethoxy)ethoxy)stilbene
  • Amyloid beta-Peptides
Funding
  • European Commission - European Medical Information Framework [115372]
Citation (ISO format)
YOUNG, Peter et al. Impact of simulated reduced injected dose on the assessment of amyloid PET scans. In: European journal of nuclear medicine and molecular imaging, 2024, vol. 51, n° 3, p. 734–748. doi: 10.1007/s00259-023-06481-0
Main files (1)
Article (Published version)
Secondary files (1)
Identifiers
Additional URL for this publicationhttps://link.springer.com/10.1007/s00259-023-06481-0
Journal ISSN1619-7070
116views
23downloads

Technical informations

Creation29/10/2023 18:04:23
First validation20/02/2024 16:13:03
Update time20/02/2024 16:13:03
Status update20/02/2024 16:13:03
Last indexation01/11/2024 08:34:28
All rights reserved by Archive ouverte UNIGE and the University of GenevaunigeBlack