Comparison of inspiratory effort with three variable-flow nasal continuous positive airway pressure devices in preterm infants: a cross-over study

Published inArchives of disease in childhood. Fetal and neonatal edition, vol. 106, no. 4, p. 404-407
Publication date2021-06-18
First online date2021-01-15
Abstract

Objective: Patient's work of breathing may vary between different neonatal nasal continuous positive airway pressure (NCPAP) devices. Therefore, we aimed to compare the inspiratory effort of three variable-flow NCPAP delivery systems used in preterm infants.

Design: Cross-over study.

Patients/setting: From June 2015 to August 2016, 20 preterm infants weighing ≤2500 g requiring NCPAP for mild respiratory distress syndrome were enrolled.

Interventions: Each patient was successively supported by three randomly assigned variable-flow NCPAP systems (MedinCNO, Infant Flow and Servo-i) for 20 min while maintaining the same continuous positive airway pressure level as the patient was on before the study period.

Main outcome measures: Patients' inspiratory effort was estimated by calculating the sum of the difference between maximal inspiratory and baseline electrical activity of the diaphragm (∆EAdi) for 30 consecutive breaths, and after normalising this obtained value for the timing of the 30 breaths.

Results: Physiological parameters (oxygen saturation measured by pulse oximetry, respiratory rate, heart beat, transcutaneous partial pressure CO2) and oxygen requirements remained identical between the three NCPAP systems. Although a wide variability in inspiratory effort could be observed, there were no statistically significant differences between the three systems for the sum of ∆EAdi for 30 breaths: CNO, 262 (±119) µV; IF, 352 (±262) µV; and SERVO-i, 286 (±126) µV, and the ∆EAdi reported on the timing of 30 breaths (sum ∆EAdi/s): CNO, 6.1 (±2.3) µV/s; IF, 7.9 (±4.9) µV/s; SERVO-i, 7.6 (±3.6) µV/s.

Conclusion: In a neonatal population of preterm infants, inspiratory effort is comparable between the three tested modern variable-flow NCPAP devices.

eng
Keywords
  • Neonatology
  • Physiology
  • Qualitative research
  • Apgar Score
  • Continuous Positive Airway Pressure / instrumentation
  • Cross-Over Studies
  • Diaphragm / physiology
  • Female
  • Heart Rate
  • Humans
  • Infant, Low Birth Weight
  • Infant, Newborn
  • Infant, Premature / physiology
  • Male
  • Oximetry
  • Oxygen / blood
  • Respiratory Distress Syndrome, Newborn / therapy
  • Respiratory Rate
Citation (ISO format)
BORDESSOULE, Alice et al. Comparison of inspiratory effort with three variable-flow nasal continuous positive airway pressure devices in preterm infants: a cross-over study. In: Archives of disease in childhood. Fetal and neonatal edition, 2021, vol. 106, n° 4, p. 404–407. doi: 10.1136/archdischild-2020-320531
Main files (1)
Article (Published version)
Secondary files (1)
Identifiers
ISSN of the journal1359-2998
22views
3downloads

Technical informations

Creation10/03/2022 1:22:06 PM
First validation08/15/2023 7:34:59 AM
Update time08/15/2023 7:34:59 AM
Status update08/15/2023 7:34:59 AM
Last indexation05/06/2024 4:51:35 PM
All rights reserved by Archive ouverte UNIGE and the University of GenevaunigeBlack