Assessing Enhanced Acoustic Absorption From Sonosensitive Perfluorocarbon Emulsion With Magnetic Resonance-Guided High-Intensity Focused Ultrasound and a Percolated Tissue-Mimicking Flow Phantom

Published inUltrasound in medicine & biology, vol. 49, no. 7, p. 1510-1517
Publication date2023-07
First online date2023-04-26
Abstract

Objective: Sonosensitive high-boiling point perfluorocarbon F8TAC18-PFOB emulsions previously exhibited thermal enhancement during focused ultrasound heating in ex vivo pig livers, kidneys and a laminar flow phantom. The main objectives of this study were to evaluate heating under turbulent conditions, observe perfusion effects, quantify heating in terms of acoustic absorption and model the experimental data.

Methods: In this study, similar perfluorocarbon emulsions were circulated at incremental concentrations of 0.07, 0.13, 0.19 and 0.25% v:v through a percolated turbulent flow phantom, more representative of the biological tissue than a laminar flow phantom. The concentrations represent the droplet content in only the perfused fluid, rather than the droplet concentration throughout the entire cross-section. The temperature was measured with magnetic resonance thermometry, during focused ultrasound sonications of 67 W, 95% duty cycle and 33 s duration. These were used in Bioheat equation simulations to investigate in silico the thermal phenomena. The temperature change was compared with the control condition by circulating de-gassed and de-ionized water through the flow phantom without droplets.

Results: With these 1.24 µm diameter droplets at 0.25% v:v, the acoustic absorption coefficient increased from 0.93 ± 0.05 at 0.0% v:v to 1.82 ± 0.22 m-1at 0.25% v:v using a 0.1 mL s-1flow rate. Without perfusion at 0.25% v:v, an increase was observed from 1.23 ± 0.07 m-1at 0.0% v:v to 1.65 ± 0.17 m-1.

Conclusion: The results further support previously reported thermal enhancement with F8TAC18-PFOB emulsion, quantified the increased absorption at small concentration intervals, illustrated that the effects can be observed in a variety of visceral tissue models and provided a method to simulate untested scenarios.

Keywords
  • Flow phantom
  • High-intensity focused ultrasound
  • Interventional radiology
  • Magnetic resonance imaging
  • Perfluorocarbons
  • Sonosensitizers
Funding
Citation (ISO format)
HOLMAN, Ryan Hawley et al. Assessing Enhanced Acoustic Absorption From Sonosensitive Perfluorocarbon Emulsion With Magnetic Resonance-Guided High-Intensity Focused Ultrasound and a Percolated Tissue-Mimicking Flow Phantom. In: Ultrasound in medicine & biology, 2023, vol. 49, n° 7, p. 1510–1517. doi: 10.1016/j.ultrasmedbio.2023.01.022
Main files (1)
Article (Published version)
Secondary files (1)
Appendix
accessLevelPublic
Identifiers
Journal ISSN0301-5629
106views
364downloads

Technical informations

Creation15/05/2023 11:53:02
First validation25/05/2023 10:35:42
Update25/05/2023 10:35:42
Status update25/05/2023 10:35:42
Last indexation01/11/2024 05:10:33
All rights reserved by Archive ouverte UNIGE and the University of GenevaunigeBlack