Analysis of the residual monomer content in milled and 3D-printed removable CAD-CAM complete dentures: an in vitro study

Published inJournal of dentistry, vol. 120, 104094
Publication date2022-05
First online date2022-03-15
Abstract

Objective: The study aimed to quantitatively evaluate the elution of methylmethacrylate from CAD-CAM manufactured removable complete dentures (RCDs) using high performance liquid chromatography (HPLC).

Methods: Thirty-two RCDs were manufactured following either the CNC-milling (Milled: n=8) or the 3D-printing (n=24) protocols. The 3D-printed dentures were further categorized into three groups based on their post-production rinsing cycles [Extended wash cycle (EWC), Standard wash cycle (SWC), and SWC with an additional Durécon coating (SWC2)]. HPLC was used to evaluate the methylmethacrylate concentrations (MMCs) eluted from the dentures in each group for different time periods (1, 2, 4, 8, and 24 hours). Mean and standard deviations were calculated for the MMCs; data was verified for normal distribution, ANOVA and post hoc tests were applied for statistical analyses (⍺=0.05).

Results: The HPLC revealed that all the denture groups recorded some amounts of MMCs, with significant differences [F (3, 31) = 23.646, p<0.0001]. The milled denture group had the highest MMCs at 24 hours when compared to the EWC (p<0.0001), SWC (p=0.001), and SWC2 (p<0.0001) denture groups. SWC had a higher MMC than EWC (p=0.032) and SWC2 (p=0.015). No differences were found in MMCs when comparing EWC and SWC2 (p=0.989).

Conclusion: Methylmethacrylate concentrations were significantly lower in 3D-printed RCDs than in milled RCDs when using the resins employed in this study. Furthermore, the MMCs can be further decreased in 3D-printed RCDs when coated with an additional thin protective layer (Durécon) by following the manufacturer-recommended rinsing protocol or when an extended isopropanol wash cycle is adopted.

eng
Keywords
  • 3D-printing
  • CAD-CAM
  • CAD-CAM milling
  • Geriatric dentistry
  • Removable complete dentures
  • Residual monomer content
  • Computer-Aided Design
  • Denture, Complete
  • Denture, Partial
  • Methylmethacrylate
  • Printing, Three-Dimensional
Citation (ISO format)
SRINIVASAN, Murali et al. Analysis of the residual monomer content in milled and 3D-printed removable CAD-CAM complete dentures: an in vitro study. In: Journal of dentistry, 2022, vol. 120, p. 104094. doi: 10.1016/j.jdent.2022.104094
Main files (1)
Article (Published version)
Identifiers
ISSN of the journal0300-5712
63views
64downloads

Technical informations

Creation09/22/2022 1:50:00 PM
First validation09/22/2022 1:50:00 PM
Update time03/16/2023 7:38:54 AM
Status update03/16/2023 7:38:52 AM
Last indexation02/01/2024 8:49:36 AM
All rights reserved by Archive ouverte UNIGE and the University of GenevaunigeBlack