A prospective evaluation of the repeatability and reliability of new steady-state pattern electroretinogram parameters

Published inJournal of Glaucoma, vol. 27, no. 12, p. 1079-1085
Publication date2018
Abstract

Purpose: The purpose of this study was to determine the repeatability and reproducibility of a new Pattern Electroretinography (PERG) device in healthy subjects. Methods: A total of 40 eyes of 20 healthy subjects (45.6 +/- 6.8 y, 70% female) were enrolled in this prospective study. Steady-state PERG were recorded using the Diopsys NOVA (Diopsys Inc., NJ) to obtain the Magnitude, MagnitudeD, MagnitudeD/Magnitude Ratio, and Signal to Noise Ratio. Subjects underwent 2 "contrast sensitivity" protocols with the same electrodes and 2 after changing electrodes, by a single examiner. Eyes were stimulated separately using high and low contrast patterns. The same examination was repeated by a different operator following the same protocol 6 months after the baseline measures. Repeatability and reproducibility of the readings were calculated using intraclass correlation coefficient (ICC). A Bland-Altman plot was used to analyze the agreement between measures. Results: ICC of Magnitude, MagnitudeD, MagnitudeD/Magnitude ratio, and SNR were 0.87, 0.81, 0.36, and 0.81, respectively between measures taken with the same electrodes (repeatability). After changing electrodes, ICC of the Magnitude, MagD, MagD/Mag ratio, and SNR were 0.79, 0.85, 0.51, and 0.83, respectively during the same session as baseline measures, and 0.87, 0.85, 0.40, and 0.76 when repeated after 6 months (reproducibility). The variability of MagD/Mag ratio increased with successive test repetitions. Conclusions: Mag, MagD, and SNR were repeatable and reproducible, whereas the MagD/Mag ratio was less repeatable and reproducible. These results suggest that the former parameters are sufficiently reliable to be used in clinical practice.

Keywords
  • Adult
  • Contrast Sensitivity/physiology
  • Electroretinography/standards
  • Female
  • Healthy Volunteers
  • Humans
  • Intraocular Pressure/physiology
  • Male
  • Middle Aged
  • Prospective Studies
  • Reproducibility of Results
  • Retinal Ganglion Cells/physiology
Affiliation entities Not a UNIGE publication
Citation (ISO format)
GILLMANN, Kevin et al. A prospective evaluation of the repeatability and reliability of new steady-state pattern electroretinogram parameters. In: Journal of Glaucoma, 2018, vol. 27, n° 12, p. 1079–1085. doi: 10.1097/IJG.0000000000001103
Main files (1)
Article (Published version)
accessLevelRestricted
Identifiers
Journal ISSN1057-0829
237views
2downloads

Technical informations

Creation13/01/2021 17:26:00
First validation13/01/2021 17:26:00
Update15/03/2023 23:52:51
Status update15/03/2023 23:52:51
Last indexation31/10/2024 21:01:48
All rights reserved by Archive ouverte UNIGE and the University of GenevaunigeBlack