UNIGE document Scientific Article
previous document  unige:101836  next document
add to browser collection
Title

Accuracy and consistency of anti-Xa activity measurement for determination of rivaroxaban plasma levels

Authors
Studt, J-D
Alberio, L
Angelillo-Scherrer, A
Asmis, L M
Korte, W
Mendez, A
Schmid, P
show hidden authors show all authors [1 - 12]
Published in Journal of Thrombosis and Haemostasis. 2017, vol. 15, no. 8, p. 1576-1583
Abstract Essentials Accurate determination of anticoagulant plasma concentration is important in clinical practice. We studied the accuracy and consistency of anti-Xa assays for rivaroxaban in a multicentre study. In a range between 50 and 200 μg L-1 , anti-Xa activity correlated well with plasma concentrations. The clinical value might be limited by overestimation and intra- and inter-individual variation.
Identifiers
PMID: 28574652
Full text
Article (Published version) (253 Kb) - document accessible for UNIGE members only Limited access to UNIGE
Structures
Research group Geneva Platelet Group (13)
Citation
(ISO format)
STUDT, J-D et al. Accuracy and consistency of anti-Xa activity measurement for determination of rivaroxaban plasma levels. In: Journal of Thrombosis and Haemostasis, 2017, vol. 15, n° 8, p. 1576-1583. https://archive-ouverte.unige.ch/unige:101836

15 hits

0 download

Update

Deposited on : 2018-01-31

Export document
Format :
Citation style :