Abstract

Fractographic analysis of clinically failed dental ceramics can provide insights as to the failure origin and related mechanisms. One anterior 6­unit all­ceramic zirconia fixed partial denture (FPD) (Cercon R) has been clinically recovered and examined using qualitative fractography. The purpose was to identify the fracture origin and to state the reasons for failure. The recovered parts of the zirconia FPD were microscopically examined to identify classic fractographic patterns such as arrest lines, hackle, twist hackle and wake hackle. The direction of crack propagation was mapped and interpreted back to the origin of failure at the interface of the occlusal–palatal tip of the core and the veneering ceramic. An inappropriate core drop design favoring localized stress concentration combined with a pore cluster in the veneering ceramic at the core tip interface were the reasons for this premature through­the­core thickness failure.

Keywords
  • Fractography
  • Failure analysis
  • Ceramic restoration
  • Zirconia
  • Fixed partial denture
  • Dental ceramic
Citation (ISO format)
LOHBAUER, Ulrich et al. Fractographic analysis of a dental zirconia framework: A case study on design issues. In: Journal of the Mechanical Behavior of Biomedical Materials, 2010, vol. 3, n° 8, p. 623–629. doi: 10.1016/j.jmbbm.2010.07.004
Main files (1)
Article (Published version)
accessLevelRestricted
Identifiers
Journal ISSN1751-6161
406views
1downloads

Technical informations

Creation10/18/2017 12:31:00 PM
First validation10/18/2017 12:31:00 PM
Update time03/15/2023 2:29:36 AM
Status update03/15/2023 2:29:36 AM
Last indexation10/31/2024 8:51:04 AM
All rights reserved by Archive ouverte UNIGE and the University of GenevaunigeBlack